Roche Diagnostics
Operator’s Manual · Version 3.1
E-75
Cedex Bio System
15 ISE troubleshooting
List of ISE flags
Out of Rng
Meaning
Slope out of range.
The slope for the specified electrode is outside the acceptable/programmed range.
Message ID
20
Priority
15
Possible cause
This flag appears with calibrations because:
o
There are problems with the ISE solutions on the instrument.
o
There are problems with the electrodes or the ISE unit was not properly serviced.
o
The ISE Reference Solution tubing is blocked or leaking.
o
The dosing from the sample probe is not accurate.
o
The preamplifier is not working properly.
Recommended actions
If only one calibration is flagged:
1. Repeat the calibration.
2. Repeat the calibration with fresh ISE Solutions 1 and 2, and check visually the
correct flow of all ISE solutions and the ISE Reference Solution.
If successive calibrations are flagged:
1. Repeat the calibration with fresh ISE Solutions 1 and 2, and check visually the
correct flow of all ISE solutions and the ISE Reference Solution.
2. Check whether the electrode expiration date has expired. Replace the electrodes, if
necessary.
3. Perform the maintenance action
Electrode Service
.
4. Perform the maintenance action
Deproteinize Probes
.
5. Check ISE Reference Solution tubing for leaks and blockages.
6. If the slopes of all electrodes are zero, replace the ISE Reference Electrode.
7. Contact your service representative.
? QC
Meaning
A result where the quality control measurement is flagged.
Message ID
111
Priority
30
Possible cause
The last QC measurement before this measurement was flagged. The dependency is
derived from the time sequence.
Recommended actions
Check QC measurement.
Summary of Contents for Cedex Bio
Page 1: ...Cedex Bio System Operator s Manual Software Version 2 1...
Page 10: ......
Page 22: ...Roche Diagnostics A 14 Operator s Manual Version 3 1 1 Safety Cedex Bio System Safety labels...
Page 130: ...Roche Diagnostics A 122 Operator s Manual Version 3 1 4 Software Cedex Bio System Buttons...
Page 131: ...5 Daily operation B 3 6 Special operations B 91 7 Configuration B 133 Operation B...
Page 132: ......
Page 303: ...8 General maintenance C 3 Maintenance C...
Page 304: ......
Page 339: ...9 Messages and alarms D 3 10 Result flags D 23 11 Troubleshooting D 39 Troubleshooting D...
Page 340: ......
Page 392: ......
Page 471: ...Glossary F 3 Index F 11 Glossary and Index F...
Page 472: ......
Page 492: ...Roche Diagnostics F 22 Operator s Manual Version 3 1 Index Cedex Bio System...
Page 493: ...Revisions G...
Page 494: ......
Page 495: ...Roche Diagnostics Operator s Manual Version 3 1 G 3 Cedex Bio System 18 Revisions Revisions...
Page 496: ...Roche Diagnostics G 4 Operator s Manual Version 3 1 18 Revisions Cedex Bio System...