13 ISE operation
Cedex Bio System
Cleaning the ISE tower off the instrument
Cleaning the ISE tower off the instrument
If performing the
Clean ISE Tower Manually
maintenance action did not lead to the
desired results, you need to remove the ISE tower to clean it.
Cleaning the ISE tower off the instrument consists of the following steps:
1. Removing the ISE tower
2. Soaking the ISE tower in ISE Deproteinizer
3. Cleaning and drying the ISE tower
4. Installing the ISE tower
5. Performing the maintenance action
Clean ISE Tower Automatically
.
Tools and materials required
o
Cotton swabs
o
ISE Deproteinizer
o
Small glass beaker
a
To clean the ISE tower
Removing the ISE tower
1
Make sure the system is in
Standby
status.
2
Shut down the system and switch off the instrument.
3
Move the transfer head to its rightmost position.
4
Open the main cover and the left service flap.
5
Remove the ISE overflow collector by pulling it upwards and turning it from side
to side as you do so.
Place the collector behind the wash station.
Make sure that you have read and understood section
The following warning messages in particular are relevant:
o
Infection by biohazardous materials
o
Injury through reagents and other working solutions
A
ISE tower
B
ISE overflow collector
C
Locking ring
Figure E-7
B
C
A
Summary of Contents for Cedex Bio
Page 1: ...Cedex Bio System Operator s Manual Software Version 2 1...
Page 10: ......
Page 22: ...Roche Diagnostics A 14 Operator s Manual Version 3 1 1 Safety Cedex Bio System Safety labels...
Page 130: ...Roche Diagnostics A 122 Operator s Manual Version 3 1 4 Software Cedex Bio System Buttons...
Page 131: ...5 Daily operation B 3 6 Special operations B 91 7 Configuration B 133 Operation B...
Page 132: ......
Page 303: ...8 General maintenance C 3 Maintenance C...
Page 304: ......
Page 339: ...9 Messages and alarms D 3 10 Result flags D 23 11 Troubleshooting D 39 Troubleshooting D...
Page 340: ......
Page 392: ......
Page 471: ...Glossary F 3 Index F 11 Glossary and Index F...
Page 472: ......
Page 492: ...Roche Diagnostics F 22 Operator s Manual Version 3 1 Index Cedex Bio System...
Page 493: ...Revisions G...
Page 494: ......
Page 495: ...Roche Diagnostics Operator s Manual Version 3 1 G 3 Cedex Bio System 18 Revisions Revisions...
Page 496: ...Roche Diagnostics G 4 Operator s Manual Version 3 1 18 Revisions Cedex Bio System...