Roche Diagnostics
E-22
Operator’s Manual · Version 3.1
13 ISE operation
Cedex Bio System
Daily operation
Replacing ISE fluid bottles
A bottle with a blue label for ISE Calibrator indirect/urine and a bottle with a red
label for ISE Reference Solution are located on the ISE unit. The procedure for
exchanging these bottles is the same.
a
To replace an on-board ISE fluid bottle
1
Follow the instructions given in
.
2
When you have finished replacing electrodes and ISE fluid bottles, press
to
proceed to the next stage in the
Prepare
wizard.
A screen is displayed that lists the maintenance actions.
Performing maintenance actions
The routine maintenance actions that are relevant for working with the ISE unit are
integrated in one single maintenance action:
Daily Prepare Actions
.
e
For general information on performing maintenance actions, see
For details on how to perform individual ISE maintenance actions, see
a
To perform maintenance actions
1
Select
Daily Prepare Actions
.
2
Press .
The maintenance definition screen is displayed.
3
Follow the instructions given on the screen.
e
4
Perform the non-routine ISE maintenance actions.
Follow the instructions given on the screen.
o
Allow at least 15 seconds for priming the ISE bottle before continuing with any
other activity, e.g. replacing the other ISE bottle.
o
After replacing an ISE fluid bottle, the
Prime ISE Reference and Calib.
mainte-
nance action needs to be performed. This action will be performed as part of the
maintenance procedure of the
Prepare
wizard.
o
After replacing an ISE fluid bottle, a main calibration needs to be performed. This
action will be performed as part of the
Daily Prepare Actions
.
When performing maintenance actions without using the Prepare wizard:
Choose
Utilities
>
Maintenance
.
Summary of Contents for Cedex Bio
Page 1: ...Cedex Bio System Operator s Manual Software Version 2 1...
Page 10: ......
Page 22: ...Roche Diagnostics A 14 Operator s Manual Version 3 1 1 Safety Cedex Bio System Safety labels...
Page 130: ...Roche Diagnostics A 122 Operator s Manual Version 3 1 4 Software Cedex Bio System Buttons...
Page 131: ...5 Daily operation B 3 6 Special operations B 91 7 Configuration B 133 Operation B...
Page 132: ......
Page 303: ...8 General maintenance C 3 Maintenance C...
Page 304: ......
Page 339: ...9 Messages and alarms D 3 10 Result flags D 23 11 Troubleshooting D 39 Troubleshooting D...
Page 340: ......
Page 392: ......
Page 471: ...Glossary F 3 Index F 11 Glossary and Index F...
Page 472: ......
Page 492: ...Roche Diagnostics F 22 Operator s Manual Version 3 1 Index Cedex Bio System...
Page 493: ...Revisions G...
Page 494: ......
Page 495: ...Roche Diagnostics Operator s Manual Version 3 1 G 3 Cedex Bio System 18 Revisions Revisions...
Page 496: ...Roche Diagnostics G 4 Operator s Manual Version 3 1 18 Revisions Cedex Bio System...