Roche Diagnostics
E-30
Operator’s Manual · Version 3.1
13 ISE operation
Cedex Bio System
Replacing ISE fluid bottles
Replacing ISE fluid bottles
ISE fluid bottles are supplied with a barcode that contains information on their
expiration date, which is monitored by the system.
Tools and materials required
o
ISE Calibrator indirect/urine or ISE Reference Solution bottle
o
Tissues
a
To replace an on-board ISE fluid bottle
1
Check that the system is in
Standby
status.
2
Choose
Overview
>
.
3
Check the status of the ISE bottles (A).
Interpreting the bottle status
screen
4
Press
Cal
or
Ref
.
A screen is displayed that shows details about the bottle status.
Roche recommend replacing ISE fluid bottles as soon as their expiration date has passed.
(Their icon on the screen turns yellow.)
For information on the stability of fluids, see their package inserts.
WARNING
Injury through working solutions
Direct contact with cleaning solutions or other working solutions may cause personal
injury.
When handling such solutions, exercise the precautions required for handling them,
observe the cautions given in the package insert, and observe the information given in the
Safety Data Sheets available for Roche Diagnostics cleaning solutions.
A
No fluid registered by an ISE fluid sensor. (Operation has stopped.)
You need to replace the bottle now.
The fluid level in the bottle is low. (Operation will proceed until one of the
sensors detects that there is no fluid.)
The fluid’s expiration date has passed.
You should replace the bottle as soon as possible.
No action is required.
Summary of Contents for Cedex Bio
Page 1: ...Cedex Bio System Operator s Manual Software Version 2 1...
Page 10: ......
Page 22: ...Roche Diagnostics A 14 Operator s Manual Version 3 1 1 Safety Cedex Bio System Safety labels...
Page 130: ...Roche Diagnostics A 122 Operator s Manual Version 3 1 4 Software Cedex Bio System Buttons...
Page 131: ...5 Daily operation B 3 6 Special operations B 91 7 Configuration B 133 Operation B...
Page 132: ......
Page 303: ...8 General maintenance C 3 Maintenance C...
Page 304: ......
Page 339: ...9 Messages and alarms D 3 10 Result flags D 23 11 Troubleshooting D 39 Troubleshooting D...
Page 340: ......
Page 392: ......
Page 471: ...Glossary F 3 Index F 11 Glossary and Index F...
Page 472: ......
Page 492: ...Roche Diagnostics F 22 Operator s Manual Version 3 1 Index Cedex Bio System...
Page 493: ...Revisions G...
Page 494: ......
Page 495: ...Roche Diagnostics Operator s Manual Version 3 1 G 3 Cedex Bio System 18 Revisions Revisions...
Page 496: ...Roche Diagnostics G 4 Operator s Manual Version 3 1 18 Revisions Cedex Bio System...