Roche Diagnostics
C-28
Operator’s Manual · Version 3.1
8 General maintenance
Cedex Bio System
Maintenance actions
Perform an air/water calibration
f
Utilities
>
Maintenance
>
Abs. Air/Water Calibration
The absorbance photometer calibration determines the air/water correction values
that are needed for result calculation. In addition, the obtained absorbance values for
an empty cuvette at each wavelength are used as a reference for the cuvette blank
check.
Operator time
Approximately 5 minutes.
Prerequisites
The system must be in
Standby
status.
Tools and materials required
m
2 cuvette segments with empty cuvettes
a
To perform an air/water calibration
1
Choose
Utilities
>
Maintenance
>
Abs. Air/Water Calibration
.
2
Press .
Your are asked to insert a cuvette segment with empty cuvettes.
3
Open the main cover.
4
If there is a segment in the cuvette ring remove it and check whether there are
used cuvettes.
If all cuvettes of this segment are empty, you can reinsert it.
5
Place a cuvette segment with empty cuvettes on the cuvette ring.
6
Press .
Your are asked to insert a second cuvette segment with empty cuvettes.
7
If there is a segment on in the cuvette ring remove it and check whether there are
used cuvettes.
If all cuvettes of this segment are empty, you can reinsert it.
8
Place the cuvette segment with empty cuvettes on the cuvette ring.
9
Close the main cover.
10
Press .
The system performs the calibration measurements. (It first measures all cuvettes
empty, pipettes water in the cuvettes, and measures them again.)
A message will inform you when the measurements are complete.
11
Scroll to display the results, if required.
12
Check the
Outlier statistics
.
Make sure that you have read and understood section
The following warning messages in particular are relevant:
o
Injury through working solutions
o
Infection by biohazardous materials
o
Personal injury or damage to the analyzer due to contact with instrument mechanism
Summary of Contents for Cedex Bio
Page 1: ...Cedex Bio System Operator s Manual Software Version 2 1...
Page 10: ......
Page 22: ...Roche Diagnostics A 14 Operator s Manual Version 3 1 1 Safety Cedex Bio System Safety labels...
Page 130: ...Roche Diagnostics A 122 Operator s Manual Version 3 1 4 Software Cedex Bio System Buttons...
Page 131: ...5 Daily operation B 3 6 Special operations B 91 7 Configuration B 133 Operation B...
Page 132: ......
Page 303: ...8 General maintenance C 3 Maintenance C...
Page 304: ......
Page 339: ...9 Messages and alarms D 3 10 Result flags D 23 11 Troubleshooting D 39 Troubleshooting D...
Page 340: ......
Page 392: ......
Page 471: ...Glossary F 3 Index F 11 Glossary and Index F...
Page 472: ......
Page 492: ...Roche Diagnostics F 22 Operator s Manual Version 3 1 Index Cedex Bio System...
Page 493: ...Revisions G...
Page 494: ......
Page 495: ...Roche Diagnostics Operator s Manual Version 3 1 G 3 Cedex Bio System 18 Revisions Revisions...
Page 496: ...Roche Diagnostics G 4 Operator s Manual Version 3 1 18 Revisions Cedex Bio System...