7 Configuration
Applications
Calculation definitions
Calculation definitions (ratio applications)
Item
Values
Comment
Factor
Used for sample and control results.
Final result = (result *
Factor
) +
Offset
The correlation factor is used to compensate for
differences in measuring methods of different
instruments. By applying a factor, the final
results of the different instruments can be
directly compared.
Offset
The offset is used to compensate for differences
in measuring methods of different instruments.
By adding the offset, the final results of the dif-
ferent instruments can be directly compared.
Valuation
No Valuation
Reference Range
Use the limits defined as
Low Limit
and
High
Limit
.
Low Used
On
,
Off
Refers to
Low Limit
.
High Used
On
,
Off
Refers to
High Limit
.
Low Limit
Results that are outside the limit are flagged.
High Limit
Results that are outside the limit are flagged.
1...3:Flag QC Fail On
,
Off
Choose
On
to ensure that all sample results are
flagged that were generated using a test whose
QC results were marked with a flag.
The initial number refers to the control that was
used for performing the QC. (See also
.)
TRL
Threshold value below which the results are no
longer displayed, instead it is indicated that the
value is smaller than the TRL value and the
<Test Rng
flag is added.
Note that you can only change this value if
TRL
Check
is activated.
Table B-15
Calculation definitions
Item
Values
Comment
Application A
through D
Up to 5 alphanumeric
characters
Short names of the installed applications.
Coefficient w
through z
000 … 999
Numeric variable for use in the formula.
Formula
Defined application
and coefficient letters.
Operands:
+, -, *, /, (, ), and space
The formula relates the defined applications
with the coefficients.
For example: (A * w + B) * A / B
Table B-16
Calculation definitions for ratio applications
Summary of Contents for Cedex Bio
Page 1: ...Cedex Bio System Operator s Manual Software Version 2 1...
Page 10: ......
Page 22: ...Roche Diagnostics A 14 Operator s Manual Version 3 1 1 Safety Cedex Bio System Safety labels...
Page 130: ...Roche Diagnostics A 122 Operator s Manual Version 3 1 4 Software Cedex Bio System Buttons...
Page 131: ...5 Daily operation B 3 6 Special operations B 91 7 Configuration B 133 Operation B...
Page 132: ......
Page 303: ...8 General maintenance C 3 Maintenance C...
Page 304: ......
Page 339: ...9 Messages and alarms D 3 10 Result flags D 23 11 Troubleshooting D 39 Troubleshooting D...
Page 340: ......
Page 392: ......
Page 471: ...Glossary F 3 Index F 11 Glossary and Index F...
Page 472: ......
Page 492: ...Roche Diagnostics F 22 Operator s Manual Version 3 1 Index Cedex Bio System...
Page 493: ...Revisions G...
Page 494: ......
Page 495: ...Roche Diagnostics Operator s Manual Version 3 1 G 3 Cedex Bio System 18 Revisions Revisions...
Page 496: ...Roche Diagnostics G 4 Operator s Manual Version 3 1 18 Revisions Cedex Bio System...