Item
Values
Comment
Sequence
No Interval
The system does not inform you about the due
status.
Use this value if you intend to perform QC as
part of standard sample testing and not as a sep-
arate task, and if QC should be handled by a host
computer.
Note that an accepted QC result is required for a
test before it can be used.
Interval Only
Due whenever the reagent interval has expired.
The system automatically informs you about
due QCs.
Interval
Number of days
(hours for ISE
applications)
Use 0 (zero) if the Sequence is set to
No Interval
.
For each of the controls:
Mat. Code
0 ... 999
Unique identification of the control material.
Numbers 801 through 999 are reserved for non
Roche controls.
You need to define this initially for every appli-
cation.
QC After Cal
On
,
Off
QC needs to be performed after each calibration
of the test.
Default QC
On
,
Off
Mark the test to have QC performed as part of
the
Default QC
function.
Default QC
is an automated process for per-
forming all QC measurements that are currently
due. This is the ideal method if you want to per-
form QC periodically during routine operation.
Table B-14
Control definitions
Summary of Contents for Cedex Bio
Page 1: ...Cedex Bio System Operator s Manual Software Version 2 1...
Page 10: ......
Page 22: ...Roche Diagnostics A 14 Operator s Manual Version 3 1 1 Safety Cedex Bio System Safety labels...
Page 130: ...Roche Diagnostics A 122 Operator s Manual Version 3 1 4 Software Cedex Bio System Buttons...
Page 131: ...5 Daily operation B 3 6 Special operations B 91 7 Configuration B 133 Operation B...
Page 132: ......
Page 303: ...8 General maintenance C 3 Maintenance C...
Page 304: ......
Page 339: ...9 Messages and alarms D 3 10 Result flags D 23 11 Troubleshooting D 39 Troubleshooting D...
Page 340: ......
Page 392: ......
Page 471: ...Glossary F 3 Index F 11 Glossary and Index F...
Page 472: ......
Page 492: ...Roche Diagnostics F 22 Operator s Manual Version 3 1 Index Cedex Bio System...
Page 493: ...Revisions G...
Page 494: ......
Page 495: ...Roche Diagnostics Operator s Manual Version 3 1 G 3 Cedex Bio System 18 Revisions Revisions...
Page 496: ...Roche Diagnostics G 4 Operator s Manual Version 3 1 18 Revisions Cedex Bio System...