6 Special operations
Cedex Bio System
Deleting QC results
Setting up your calibration schedule
The
Cedex Bio
instrument provides convenient software-supported procedures for
the following major calibration scenarios:
o
Performing all calibrations that are currently due
o
Performing all calibrations that fall due within a configurable forecast period
o
Performing individual calibrations
Performing calibration as part of the Prepare wizard includes performing all
calibrations that are due and that will fall due during the forecast period.
The following table lists the configurable definitions and describes their effect.
Deleting QC results
Accepted QC results are normally deleted from the
QC Status
list as part of the daily
end of shift activities, at the same time, results that were generated before the first day
of the previous month are deleted from the
QC History
list.
This section describes how to proceed if you need to delete QC results outside the
normal daily routine.
e
For deleting QC results at the end of a shift, see
Definition item
Configuration path
Effect
Forecast Hours
Utilities
>
Configuration
>
Calibration
Period of time, starting with the current
date, within which calibration falls due.
Defining a forecast period (value other than
zero) allows to concentrate calibration per-
formance to specific days.
Cal Auto Accept
Utilities
>
Configuration
>
Result
Handling
If on, results that are not flagged are auto-
matically accepted.
Results with flags that are marked in a pre-
defined list of acceptable flags would also be
automatically accepted.
Cal Acceptable
Flags
Utilities
>
Configuration
>
Result
Handling
List of calibration relevant flags. Flags that
are marked are automatically accepted if
Cal Auto Accept
is on.
Table B-10
Calibration definitions and their effect
o
QC results are displayed in the
QC Status
list as soon as they are generated. A copy of
the results is also placed in the
QC History
list. This list keeps the results of the previ-
ous and the current months.
o
Deleting QC results, deletes the associated orders as well.
Summary of Contents for Cedex Bio
Page 1: ...Cedex Bio System Operator s Manual Software Version 2 1...
Page 10: ......
Page 22: ...Roche Diagnostics A 14 Operator s Manual Version 3 1 1 Safety Cedex Bio System Safety labels...
Page 130: ...Roche Diagnostics A 122 Operator s Manual Version 3 1 4 Software Cedex Bio System Buttons...
Page 131: ...5 Daily operation B 3 6 Special operations B 91 7 Configuration B 133 Operation B...
Page 132: ......
Page 303: ...8 General maintenance C 3 Maintenance C...
Page 304: ......
Page 339: ...9 Messages and alarms D 3 10 Result flags D 23 11 Troubleshooting D 39 Troubleshooting D...
Page 340: ......
Page 392: ......
Page 471: ...Glossary F 3 Index F 11 Glossary and Index F...
Page 472: ......
Page 492: ...Roche Diagnostics F 22 Operator s Manual Version 3 1 Index Cedex Bio System...
Page 493: ...Revisions G...
Page 494: ......
Page 495: ...Roche Diagnostics Operator s Manual Version 3 1 G 3 Cedex Bio System 18 Revisions Revisions...
Page 496: ...Roche Diagnostics G 4 Operator s Manual Version 3 1 18 Revisions Cedex Bio System...