5 Daily operation
Cedex Bio System
Finishing the shift
Cleaning up the database
The
Cedex Bio
instrument can store orders and results data for one working day. It is
therefore necessary to delete results and orders to ensure that there is enough space
on the system for the next shift.
a
To clean up the database
1
Press .
The following results will be deleted:
o
Accepted sample results.
o
QC results that were accepted are deleted from the QC Status list (they remain
in the QC history).
o
QC results that were generated before the first day of the previous month are
deleted from the QC history.
o
Obsolete calibration results, provided there are more than five obsolete results
for the same test.
(Results become obsolete if the empty set was removed or if the set was
removed more than 30 days ago.)
2
Press
to confirm the deletion.
The results are deleted.
3
Press
to proceed to the next stage in the
End Shift
wizard.
A screen for handling cuvettes is displayed.
By deleting the sample results, you delete the corresponding orders as well.
When deleting results outside the End Shift wizard:
You can delete accepted sample results, calibrations, orders, and accepted and ignored
QC results by pressing
in the respective result lists.
e
e
e
e
To delete QC results from the QC Status list
e
To delete QC results from the QC History
Maintenance action Cleanup Database
The maintenance action
Cleanup Database
is intended for situations where the system
does not work efficiently any more, it is not intended to clear up the data at the end of a
shift.
Summary of Contents for Cedex Bio
Page 1: ...Cedex Bio System Operator s Manual Software Version 2 1...
Page 10: ......
Page 22: ...Roche Diagnostics A 14 Operator s Manual Version 3 1 1 Safety Cedex Bio System Safety labels...
Page 130: ...Roche Diagnostics A 122 Operator s Manual Version 3 1 4 Software Cedex Bio System Buttons...
Page 131: ...5 Daily operation B 3 6 Special operations B 91 7 Configuration B 133 Operation B...
Page 132: ......
Page 303: ...8 General maintenance C 3 Maintenance C...
Page 304: ......
Page 339: ...9 Messages and alarms D 3 10 Result flags D 23 11 Troubleshooting D 39 Troubleshooting D...
Page 340: ......
Page 392: ......
Page 471: ...Glossary F 3 Index F 11 Glossary and Index F...
Page 472: ......
Page 492: ...Roche Diagnostics F 22 Operator s Manual Version 3 1 Index Cedex Bio System...
Page 493: ...Revisions G...
Page 494: ......
Page 495: ...Roche Diagnostics Operator s Manual Version 3 1 G 3 Cedex Bio System 18 Revisions Revisions...
Page 496: ...Roche Diagnostics G 4 Operator s Manual Version 3 1 18 Revisions Cedex Bio System...