Roche Diagnostics
Operator’s Manual · Version 3.1
B-51
Cedex Bio System
5 Daily operation
Validating sample results
Validating sample results
The
Cedex Bio
instrument provides several aids for validating results:
o
In the results list, results that fall outside predefined technical ranges are flagged.
o
You can display detailed result information that allows you to make a considered
decision.
o
Non flagged results can be accepted automatically.
o
You can print the results.
o
You can export the results and process them on an external computer.
The following ways of dealing with results are available:
o
Accept the result.
o
Re-perform the identical test (
Repeat
).
o
Re-perform the test using a predefined different dilution (
Rerun
).
Results must be accepted before they can be transmitted to a host or before they can
be automatically printed.
The effect of flagged results depends on the configuration (
Utilities
>
Configuration
>
Result Handling
). The following table shows how.
e
Editing the acceptable flags list
There are two main approaches to validating results:
o
To validate sample results of an order whose sample is still on board, press the
tube button on the
Overview
tab.
o
In all other situations, choose
Workplace
>
Result Review
.
Results must be accepted before they can be transmitted to a host or before they can be
automatically printed.
Ratio results cannot be manually accepted. They are automatically accepted if their con-
stituent results are accepted.
Configuration setting
Effect
Sample Auto Accept
:
On
Results that do not contain a flag are auto-
matically accepted. Results with flags that are
marked in a predefined list of flags that
should be ignored are accepted as well.
Sample Auto Accept
:
Off
All results need to be manually accepted.
Sample Accept Flags
Flags that are marked in this list will be
ignored by the system.
Table B-4
Sample flag configuration and its effect
Summary of Contents for Cedex Bio
Page 1: ...Cedex Bio System Operator s Manual Software Version 2 1...
Page 10: ......
Page 22: ...Roche Diagnostics A 14 Operator s Manual Version 3 1 1 Safety Cedex Bio System Safety labels...
Page 130: ...Roche Diagnostics A 122 Operator s Manual Version 3 1 4 Software Cedex Bio System Buttons...
Page 131: ...5 Daily operation B 3 6 Special operations B 91 7 Configuration B 133 Operation B...
Page 132: ......
Page 303: ...8 General maintenance C 3 Maintenance C...
Page 304: ......
Page 339: ...9 Messages and alarms D 3 10 Result flags D 23 11 Troubleshooting D 39 Troubleshooting D...
Page 340: ......
Page 392: ......
Page 471: ...Glossary F 3 Index F 11 Glossary and Index F...
Page 472: ......
Page 492: ...Roche Diagnostics F 22 Operator s Manual Version 3 1 Index Cedex Bio System...
Page 493: ...Revisions G...
Page 494: ......
Page 495: ...Roche Diagnostics Operator s Manual Version 3 1 G 3 Cedex Bio System 18 Revisions Revisions...
Page 496: ...Roche Diagnostics G 4 Operator s Manual Version 3 1 18 Revisions Cedex Bio System...