10
• Maintenance and Care
144
10.6 Deleting patient data
This feature allows you to delete all patient-related data
(patient lists and results). This may become necessary to
comply with data protection and privacy policies when
sending a meter back to the manufacturer or passing it
on to another facility.
To ensure deleted data is not transferred back through an
unwanted automatic synchronization with the DMS,
remove the battery pack immediately after deleting
patient data.
Setup Menu
Date / Time
QC
Test Display
Password
Test Ranges
Reagents
Operator ID
Patient ID
Unlock
Delete Pat.
12:48 pm
09/18/13
Confi rm Deletion
D-201: Delete Data:
Are you sure you want to
delete all patients and all
results from this meter?
This operation cannot be
undone.
Setup Menu
Date / Time
QC
Test Display
Password
Test Ranges
Reagents
Operator ID
Patient ID
Unlock
Delete Pat.
12:48 pm
09/18/13
Be aware that the function
Delete Patient
irrevocably
deletes all data once you touch
to confirm. The
system will delete any patient results regardless of
whether they have been synchronized with a data
management system or not. Therefore, always ensure
that you have downloaded all relevant patient results
before you use this function.
Summary of Contents for ACCU-CHEK Inform II
Page 1: ...Accu Chek Inform II BLOOD GLUCOSE MONITORING SYSTEM Operator s Manual ...
Page 3: ...0 4807839001 05 2013 09 EN Operator s Manual Version 5 0 A ACCU CCU CHEK HEK Inform II System ...
Page 6: ...6 This page intentionally left blank ...
Page 12: ...12 This page intentionally left blank ...
Page 104: ...8 Proficiency Testing 104 This page intentionally left blank ...
Page 160: ...12 General Product Information 160 This page intentionally left blank ...
Page 206: ...D Supplement for Observed Test Sequence 206 This page intentionally left blank ...
Page 210: ...Index 210 This page intentionally left blank ...
Page 211: ...211 ...