pH/ORP-5500 Series 

pH/ORP Transmitting Controller 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Instruction Manual 

 

Summary of Contents for pH/ORP-5500 Series

Page 1: ...pH ORP 5500 Series pH ORP Transmitting Controller Instruction Manual ...

Page 2: ...ntenance work for free or changes it 2 pH and ORP probes are consumer goods which without quality guarantee Because the lifespan is based on the medium and working conditions 3 Manufacturer offers the maintenance service for whole life of the sold meters 4 If the damage of the meter is caused by the following reasons it is out of the maintenance service A The meter is burned or foundered caused by...

Page 3: ... LAYOUT 5 3 2 ELECTRICAL CONNECTION LAYOUT 5 Ⅳ OPERATE SETTING 6 4 1 PANEL AND MAIN INTERFACE DISPLAY 6 4 2 KEY BOARD FUNCTION 7 4 3 SETTING INTERFACE 7 4 4 PROBE CALIBRATION 8 4 4 1 CALIBRATION 8 ⅤINTRUMENT MATING PROBE INSTALLATION GUIDE 9 5 1 PROBE OUTLINE DIMENSION 9 5 2 PROBE INSTALLATION 10 5 3 PROBE INSTALLATION REQUEST 11 Ⅵ MAINTAINING 12 6 1 PROBE MAINTAINING 12 6 2 PROBE USING ATTENTION ...

Page 4: ... probe easy change one time wiring permanent using see the pH ORP1220Aseries instruction 10 Industrial standard terminals EMC electromagnetic compatibility measurement platform design and development series grouping 11 Satisfy world electric power supply standard DC AC power input nonpolarity connection site operation will not worry about wrong connect 12 SMT Automatically Surface Mount Technology...

Page 5: ...0 4 01 Temp compensation error pH 0 05 Entire instrument error pH 0 1 pH 0 05 Entire instrument repeatability error ORP 2mV Compensation range Temp 0 50 Temp 0 50 Working environment Humidity 85 RH Cable length 20m standard 10m Control output double relay single contact ON OFF hysteresis can be set 3A max AC220V AC110V 3A max DC24V Contact capacity Power consumption 2 5W Channels One channel Isola...

Page 6: ...nel mounted fast installation clamp Dimension 96 mm 96mm 103 mm H W D hole dimension 91mm 91mm Ⅱ INSTRUMENT DIMENSION AND TERMINAL DEFINITION 2 1 Instrument dimension Shell front view Shell top view Shell side view 3 ...

Page 7: ...otection terminal Hi High limit alarm control terminal blocks Low Low limit alarm control terminal blocks T T T external feed under transmitting mode T I public terminal under Transmitter Instrument Mode I I I mA output terminal under Instrument mode INPUT Connect pH ORP measurement electrode Clear line REF Connect pH ORP reference electrode shielding CELL T Connect temp probe collect terminal red...

Page 8: ...ERENCE DIAGRAM 3 1 Milliampere transmitting connection layout Instrument mode current loop wiring Transmitter mode current loop wiring 3 2 Electrical connection layout Relay ON OFF contact component control wiring diagram 5 ...

Page 9: ...3 3 Control Mode alarm high low limit setting identification Units show identification pH show identification mA migrate setting identification a Ⅳ OPERATE SETTING 4 1 Panel and main interface display 6 ...

Page 10: ...save enter into next parameter setting 1 Function selection 4mAflashing operate select Add key set 4mA 4mA corresponding pH setting 2 corresponding pH value press enter key save enter into next parameter setting the same below 20mA corresponding pH setting 20mAflashing set 20mA corresponding pH value confirm 3 and save enter into next parameter setting High limit alarm pH setting Hi flashing setti...

Page 11: ... seconds enter into buffer solution calibration interface input current buffer solution tempreture press enter key save enter into buffer solution calibration interface 2 10 00 pH flashing indicate enter into buffer solution calibrate selection press enter key start calibration 10 00 flashing press return key skip it enter into next calibration menu 3 Put the cleaned probe into selected buffer sol...

Page 12: ...ation Under pH measurement main interface press select 3s enter into manul calibrate interface input the relative mA value of buffer solution with mating probe press save return main interface NOTE input wrong LCD screen display under any calibrating spot mV value Press enter key to continue calibration or press return key cancel the calibration and enter into next calibrate spot ⅤINTRUMENT MATING...

Page 13: ...on pH ORP Flow device P34A installation pH ORP Flow device P34B installation Temp probe pipeline installation pH ORP Floating bed style installation pH ORP Scaffolding installation P16 pH ORP reaction pool installation 10 ...

Page 14: ...eline pure water measurement value will unstability because the water is weak electrolyte sensitive glass film glass bubble and salt bridge present discontinuity open circuit result the measure value unstability 4 When direct install the pipeline to measure pure water high concentration KCL of pH probe and pure water form huge concentration difference trough liquid abutment spread into the water l...

Page 15: ...en use clear water washing up 7 washing cleaned probe suggested immerse into 3 0mol L KCL solution hydration 6 hours then use it 8 When calibrate pH probe and instrument the buffer solution must processing under 25 6 2 Probe using attention 1 pH ORP probe is consumption type probe it will ineffective after long time storage so suggested to buy to use do not storage 2 Prohibit sensitive glass film ...

Page 16: ...er solution to calibrate A Probe fault B Cable damage Measurement data large deviation B change the probe which can not calibrate C Installation dead area C analysis and improve measure point or using flow device D Instrument Setting error D resetting the instrument parameter A Using single calibrating instrument inspect the instrument 4 20mA A Recelive cell unreapear B Change signal loop cable le...

Page 17: ... General probe fission probe Stiring reaction kettle and material tank Fold immersed install assembly General probe fission probe Dirt waste water pool aeration basin General probe fission probe Open channel stream aquicuture Floating bed component High temp sterilization glass probe Biological Pressure sheath components pharmaceutical food fermentation Noble metal protective screening General pro...