CMM96-MD
Manual de Usuario
User´s Manual
92.0
96.0
96.0
81.5
5.3
+0,8
Agujero de panel
Panel cut-out
AL-1
AL-2
AL-3
AL-4
RMS
V
Vavg
A
Md
C-h
Aavg
Hz
L-1
L-2
L-3
N
Holding piece
Brida de sujección
92.0
+0,8
Dimensiones: 96 x 96 x 81,5
Peso:
350 gr.
Material caja: ABS autoextinguible
Color caja:
Gris Antracita
Frontal:
IP54 (IP65 opcional)
Dimensions: 96 x 96 x 81.5
Weight:
350 gr.
Case material:Self-exting ABS
Case colour: Anthracite grey
Frontal:
IP54 (IP65 optional)
Dimensiones y datos mecánicos /
Dimensions & Mechanical Data
V
V
V
V
V
V
I
I
I
Vavg
Aavg
Hz
Unb-V ; Unb-C
MD I1 ; MD I2 ; MD I3
Memorización
4 alarmas programables.
1 -
n
2 -
n
3 -
n
1 -
2
2 -
3
3 -
1
1
2
3
Medida en verdadero valor eficaz de las tensiones simples o compuestas, asi como de las
corrientes de fase. Tensión y corriente promedio de las tres fases (average). Frecuencia, con
una resolución de 0,1Hz y en el rango de 20 a 600Hz.
Desequilibrio de las tensiones y de las corrientes. Este valor es muy importante si desean
realizarse protecciones de sistemas susceptibles a los desequilibrios. Asismismo es muy útil a
la hora de hacer mantenimientos preventivos.
Tres Maxímetros de Corriente. El CMM96 MD dispone de tres maxímetros de corriente, es
posible seleccionar cualquier tiempo de integración en el rango de 5 a 60 minutos.
de los valores pico y valle de todos los parámetros asi como los tres cuenta
horas. Estos quedarán guardados aún con pérdida de la alimentación auxiliar. El CMM96
utiliza un sistema de memoria, de última generación, que no precisa de bateria externa, con lo
que no está sujeto a los problemas de mantenimiento tipicos de otros sistemas.
Es posible, por ejemplo, programar el primer relé para tener alarma
cuando cualquiera de las tres tensiones sean superiores a 250V o inferiores a 190V;el
segundo cuando cualquier corriente instantánea sea superior a 500A durante más de 30 seg,
el tercero cuando el desequilibrio en corriente sea superior al 30%; finalmente el cuarto relé
nos dará alarma cuando cualquier corriente integrada (maxímetro) sea superior a 250A
.
Seguridad y Fiabilidad
.Los instrumentos de la serie CMM96 han sido ensayados en
laboratorio, de acuerdo a las últimas normativas y directivas, y contrastados en ambientes
industriales, superando con éxito las pruebas más severas de condiciones ambientales, ruidos
eléctricos conducidos, perturbaciones electromagnéticas radiadas, vibraciones, etc.
Funciones y características principales
V
V
V
V
V
V
I
I
I
Vavg
Aavg
Hz
Unb-V ; Unb-C
MD I1 ; MD I2 ; MD I3
Memory Storage
4 Programmables Alarms.
1 -
n
2 -
n
3 -
n
1 -
2
2 -
3
3 -
1
1
2
3
True RMS measurement of the phase to phase or phase to neutral voltage, and phase´s
currents. Three phase voltage and current average. Frequency, with 0.1 Hz resolution and in
the 20 to 600Hz range.
Voltage and Current Unbalance computation. This value is very important if we want to protect
electrical or electronic equipment . Also is very useful in order to do predictive maintenance in
electrical motors.
Three Current Maximum Demand. The CMM96 MD has three current maximum demand and is
possible to select any integration time between the 5 to 60 minutes range.
of the all peak and valley value measures, and the Counter hours. The
CMM96 CH use and advanced generation memory technology, thus avoiding the value loss in
case of a power supply lack and annulling the inconveniences of using of a battery supplied
systems.
Is possible, for example, to program the first relay for have an
alarm when any voltage (Phase 1, 2 or 3) became bigger than 250V or lower than 190V; the
second when any instantaneous current became bigger to 500A during more than 30 seconds;
the third when the current unbalance achieve the 30 %; and finally the fourth relay will be
connected when any integrated current became bigger to 250A.
.
Security and Reliability.
Instruments of the CMM96 Series have been tested in laboratories
and check at industrial environments, and have successfully passed the most rigorous tests
involving ambient conditions, electrical noises, electromagnetic disturbances, vibrations, etc....
Functions and Main Characteristics
Características Técnicas /
Technical Data
Tensión y tolerancia
Margen de Frecuencia
Consumo
Consumo con tarjeta relés
Tensión Nominal
Frecuencia
Corriente Nominal
Sobrecarga Permanente
Consumo circuito Corriente
Tensión y Corriente
Frecuencia
Cuenta horas
Display
Color
Ciclo de Presentación
Leds Auxiliares
Temperatura Almacenamiento
Temperatura de Uso
230 / 400V -10/+15%
35...450Hz
5VA
7VA
400V F-N, 565 F-F
35...450Hz
5A
1.2 In
0.6VA
Cl 0.5 +/-2dig
+/- 0.01Hz
+/- 1h/año (
3 x 4 digits
Red, High Efficiency
1 x seg
10
-40...+70ºC
-10...+65ºC
year)
1 x sec
Voltage and Tolerance
Frequency band
Consumption
Consumption with relays card
Nominal Voltage
Frequency band
Nominal Current
Continuous Overload
Current circuit consumption
Voltage and Current
Frequency
Count-Hour
Display
Colour
Display rate
Auxiliar Leds
Storage Temperature
Operation Tempearture
Normas de Diseño /
Design Standards
IEC 1010, IEC 348, IEC 664, IEC 801, EN 50081-2, EN 50082-2
Alimentación Auxiliar /
Auxiliary Supply
Circuito de Medida /
Measuring Circuit
Precisión /
Accuracy
Display
Condiciones Ambientales /
Environmental conditions
Opcionales /
Optionals
Alimentación Auxiliar
/ Auxiliary Supply
9....18V dc, 18...36V dc, 36...72V dc, or 90...140Vdc.
CMM96 con entradas de corriente aisladas /
Protector Frontal IP65 /
Tarjeta 2 ó4 relés con contacto conmutado
Tarjeta 2 ó 4 relés con contacto con Control externo cuenta horas 1
CMM96 with isolated inputs current
Frontal Protector IP65
/ 2 or 4 Relays change over contact board
/
2 or 4 Relays change over contact board + External Control Count-hour 1
PN: 430029004 REV A