35-3001-05-XX CO
2
Sample-Draw Detector • 21
5. Disconnect the sample tubing from the inlet line.
6. Unscrew the regulator from the carbon dioxide calibration cylinder.
Returning to Normal Operation
1. Wait approximately one minute to allow the carbon dioxide reading to stabilize.
2. Follow the instructions in the controller’s operator’s manual to exit the calibration mode.
3. Store the components of the calibration kit in a safe and convenient place.
Parts List
Table 2 lists replacement parts and accessories for the carbon dioxide sample-draw detector.
Table 2: Parts List
Part Number
Description
06-1248RK
Sample tubing, 3/16” ID x 5/16” OD, specify length
06-1248RK-03
Sample tubing, 3/16” ID x 5/16” OD, 3 feet (for calibration kit)
07-0110RK
Gasket for IR flow block
17-2670
Inlet fitting O-ring
17-2671
Inlet fitting front ferrule
17-2672
Inlet fitting back ferrule and tube nut
30-1016RK
Pump
33-0165RK
Hydrophobic filter
33-0167RK
Particle filter
61-5040RK-02
Infrared carbon dioxide sensor, 0 - 5,000 ppm
range
61-5040RK-03
Infrared carbon dioxide sensor, 0 - 5% vol
range
61-5040RK-05
Infrared carbon dioxide sensor, 0 - 50% vol
range
61-5040RK-10
Infrared carbon dioxide sensor, 0 - 100% vol
range
71-0413
Operator’s Manual, 35-3001-05-XX Sample-Draw Detector
81-0070RK-01
Calibration cylinder, 2000 ppm CO
2
in nitrogen, 34 liter steel
81-0070RK-03
Calibration cylinder, 2000 ppm CO
2
in nitrogen, 103 liter
81-0072RK-01
Calibration cylinder, 2.5% CO
2
in nitrogen, 34 liter steel
81-0072RK-03
Calibration cylinder, 2.5% CO
2
in nitrogen, 103 liter
81-0073RK-01
Calibration cylinder, 15% CO
2
in nitrogen, 34 liter steel
81-0073RK-03
Calibration cylinder, 15% CO
2
in nitrogen, 103 liter
81-0078RK-01
Calibration cylinder, 100% nitrogen, 34 liter steel