RKI 35-3001-05 Series Operator'S Manual Download Page 14

 

14 • 35-3001-05-XX CO

2

 Sample-Draw Detector

Start Up

This section describes procedures to start up the sample-draw detector and place the sample-draw detector 
into normal operation.

Introducing Incoming Power

1. Complete the installation procedures described earlier in this manual.

2. Verify that the wiring is correct and secure. Refer to the controller operator’s manual for connections at 

the controller.

3. Turn on or plug in the power to the controller, then turn on the controller.
4. Verify that the sample-draw detector’s Pilot LED is on.
5. Verify that the controller is on and operating properly. Refer to the controller operator’s manual.
6. Verify that the flowmeter indicates a flow rate of approximately 1.0 SCFH. If necessary, use the sensor 

flow control valve to adjust the flow rate. Turn the valve’s knob counterclockwise to increase the flow 
and clockwise to decrease the flow.

NOTE:

The following step tests for leaks in the sample line. This test will cause a low flow condition at 
the sample-draw detector and a fail condition at the controller. Be sure to put the controller into its 
calibration program or disable external alarms before performing this test.

7. Verify that the incoming sample line is not leaking. To test the sample line, press and hold the reset 

switch and plug the open end of the sample line with your thumb. If the flowmeter ball drops to the 
bottom of the flowmeter, the incoming sample line is not leaking.

8. Remove your thumb from the sample line, release the reset switch, and verify the flowmeter returns to a 

normal flow rate.

9. Enable alarms or place the controller in normal operation.

CAUTION:

Allow the sample-draw detector to warm up for 5 minutes before you continue with the next 
section, “Setting the Zero Reading.”

Setting the Zero Reading

Since there is a background of CO

2

 in air of typically 300 - 600 ppm (0.03 - 0.06% volume), it is necessary 

to use a calibration kit with a 100% nitrogen cylinder to set the zero signal for the 
0 - 5,000 ppm and 0 - 5% volume versions. Fresh air can be used to set the zero signal for the 
0 - 50% and 0 - 100% volume versions.

1. Screw the demand flow regulator into the 100% nitrogen calibration cylinder.

2. Connect the calibration tubing from the demand flow regulator to the inlet fitting. Gas will begin to 

flow.

3. Allow the sample-draw detector to draw sample for one minute.

4. Verify a reading of 0% vol on the controller display screen for the applicable channel.

If the display reading is 0% vol, start up is complete. The sample-draw detector is in normal operation. 
If the display reading is not 0% vol, continue with step 5.

5. Perform a zero operation at the controller. See “Setting the Zero Reading” on page 20 and the controller 

operator’s manual for instructions to perform a zero operation.

6. Remove the calibration tubing from the inlet fitting, then reconnect the sample tubing to the inlet fitting.

www.

.com

[email protected]

1.800.561.8187

Summary of Contents for 35-3001-05 Series

Page 1: ...35 3001 05 XX Carbon Dioxide Sample Draw Detector Operator s Manual Part Number 71 0413 Revision A Released 6 20 19 www com information itm com 1 800 561 8187...

Page 2: ...aintenance of the detector is essential for proper operation and correct readings Please calibrate and maintain this detector regularly Frequency of calibration depends upon the type of use you have a...

Page 3: ...epair costs transportation costs or contingent expenses incurred without our prior approval THIS WARRANTY IS EXPRESSLY IN LIEU OF ANY AND ALL OTHER WARRANTIES AND REPRESENTATIONS EXPRESSED OR IMPLIED...

Page 4: ...Controller 12 Start Up 14 Introducing Incoming Power 14 Setting the Zero Reading 14 Maintenance 15 Preventive Maintenance 15 Troubleshooting 15 Replacing Components of the Carbon Dioxide Sample Draw D...

Page 5: ...ury Be sure to maintain and periodically calibrate the 35 3001 05 XX as described in this manual Table 1 Specifications Target Gas Carbon Dioxide CO2 Detection Range 35 3001 05 02 0 5 000 ppm parts pe...

Page 6: ...sor P CB B G Detector Amp Terminal Strip Factory W ired AMP AMP 2 S Sensor Flow Control V alve Fail LED Pilot LED Pump Reset Switch Flowmeter Pressure Switch Adjustment Screw behind P CB R B G Interco...

Page 7: ...rigid tubing See the Installation section on page 10 to connect tubing to the sample fittings Particle Filter A particle filter with a tubing stub on one end is shipped with the instrument but it is n...

Page 8: ...onnector to connect the pump to the main circuit board NOTE The pump is factory wired to the main circuit board See Installation on page 10 for all wiring procedures related to the sample draw detecto...

Page 9: ...tion and start the pump again press and hold the pump reset switch for about 2 seconds then release Hydrophobic Filter The hydrophobic filter is located toward the bottom left of the main circuit boar...

Page 10: ...en you select the mounting site Is there enough room to open the housing door and make wiring and tubing connections at the bottom of the housing Make sure there is sufficient room to perform start up...

Page 11: ...the sampling area 1 4 O D rigid polypropylene Teflon or flexible polyurethane tubing may be used RKI Instruments Inc recommends using either 1 4 O D x 1 8 I D or 1 4 O D x 0 170 I D tubing based on yo...

Page 12: ...ing so the ferrules clamp on the tubing If the tube nut is removed see Figure 5 for the arrangement of the fitting components 4 Route the opposite end of the tubing to an open area where the sample ca...

Page 13: ...r 9 If shielded cable is used connect the cable s drain wire to an available chassis earth ground at the controller RKI controllers typically have a ground stud that can be used to ground the cable s...

Page 14: ...er ball drops to the bottom of the flowmeter the incoming sample line is not leaking 8 Remove your thumb from the sample line release the reset switch and verify the flowmeter returns to a normal flow...

Page 15: ...not show a reading in fresh air Investigate significant changes in the display reading Biannual Calibration Calibrate the sample draw detector as described in Calibration on page 20 Troubleshooting T...

Page 16: ...e line for obstructions or kinks 5 If the calibration response difficulties continue replace the sensor as described later in this section 6 If the calibration response difficulties continue contact R...

Page 17: ...e the housing door Replacing the Particle Filter 1 Turn off the controller 2 Turn off power to the controller 3 If the particle filter is installed at the inlet fitting 1 Disconnect the tubing routed...

Page 18: ...set the flow again Repeat this step until you are able to adjust the flow to 0 6 SCFH NOTE The pressure switch adjustment screw is accessible through a circular cutout in the flowmeter circuit board...

Page 19: ...t Be careful not to lose the O ring that may come out with the tubing stub If you do lose the O ring or if it is damaged see Parts List on page 21 for the spare part number 3 Push the tubing stub up a...

Page 20: ...l for entering calibration mode NOTE If you can verify a fresh air environment it is not necessary to use the zero air calibration cylinder to set the zero reading 2 Screw the regulator into a zero ai...

Page 21: ...O ring 17 2671 Inlet fitting front ferrule 17 2672 Inlet fitting back ferrule and tube nut 30 1016RK Pump 33 0165RK Hydrophobic filter 33 0167RK Particle filter 61 5040RK 02 Infrared carbon dioxide s...

Page 22: ...r demand flow for 34 liter aluminum 58 liter and 103 liter cali bration cylinders cylinders with internal threads 81 1055RK Regulator demand flow for 17 liter and 34 liter steel calibration cylinders...

Reviews: