Alarm Check • 100
should return to its unused state at the top of the SDM-6000.
24 . Remove the GX-6000 from the SDM-6000.
25 . Repeat step 21 - step 24 for any additional GX-6000s.
Alarm Check
To perform an alarm check on a GX-6000:
1 . Turn on the SDM-6000 system as described in “Turning on a SDM-6000
2 . Confirm that you have an SD card installed in the system.
NOTE:
If an SD card is not installed, then test data and instrument data logger
data downloaded manually or automatically cannot be saved.
3 . If you have a multi-station system and a slave station is turned on but you do
not want it to be included in the testing, deselect the Active box in the Unit Tab.
4 . If necessary, press the Function Tab in the upper left corner to get the Function
Tab to appear.
5 . Press the Alarm Check button to initiate an alarm check.
If you wish to cancel the alarm check, press the Alarm Check Cancel button in
the Function Tab and press Yes in the window that appears.
6 . If there is no SD card installed, a warning screen will appear. If no SD card is
installed, no test data will be saved. Press Yes or No to continue with the alarm
check.
7 . The buzzer will sound, the LEDs will turn on for a few seconds, the vibrator will
vibrate, and the SDM-6000 will determine if these actions were completed
successfully.
8 . Once the alarm check is over, the Function Tab will display an Alarm Check
End button. To view the test results, press the Unit button to display the Unit