background image

 

12 • Maintenance

Model GX-82 Instruction Manual

Maintenance

Batteries

1. Check the battery voltage periodically by pressing the BATT button. Replace the 

batteries before the voltage drops to the operational limit (see “Alarms” on page 7).

WARNING: Take the GX-82 to a non-hazardous location before replacing the batteries.

2. To replace the batteries, slide the battery compartment cover off the instrument case. 

Remove the batteries and verify that the battery compartment and electrical contacts 
are clean. If necessary, use a soft wire brush to gently clean the compartment and 
contacts. Insert fresh batteries (alkaline or fully charged Ni-Cd) according to the 
polarity (+/-) marking, then replace the cover.

3. Bias Current Discharge

The batteries continuously supply a small current to maintain the H

2

S and CO 

sensors, even when the GX-82 is off (see Sensor Maintenance). This current drain is 
minimal but will result in a normal discharge of the batteries over a period of several 
weeks.

NOTE: If the batteries are fully discharged before replacement, allow up to 1/2 hour for 

the H

2

S and CO circuits to show a normal response.

CO Filter

CO sensors are equipped with an activated carbon filter that removes H

2

S and most 

hydrocarbons to limit interference with the CO measurement.

To install the filter, slip it over the face of the CO sensor. Make sure the filter fits snugly 
and is seated properly. Replace the filter when CO readings become suspect (for example, 
the CO circuit calibrates properly but shows response in a known CO-free environment).

Sensor Maintenance

Electrochemical sensors (O

2

, H

2

S, CO) gradually deteriorate, regardless of use, and 

require periodic replacement. Combustibles sensor life is generally related to usage, but 
certain environmental factors may affect duration.

The GX-82 sensors are easy to replace, but only the combustibles sensor contains user-
serviceable components. If a sensor requires replacement, call RKI or your local 
distributor. All sensors are warranted usable for one year from the date of shipment. 
Sensors that fail within the warranty period will be replaced at no charge.

Combustibles sensor

Replace the combustibles sensor or filaments when:

the combustibles circuit cannot be calibrated correctly

the COMB (%LEL) display does not show 0 immediately after the start-up sequence, 
and cannot be set to zero by the Auto Zero command

Summary of Contents for GX-82

Page 1: ...RKI Instruments Inc 33248 Central Ave Union City CA 94587 510 441 5656 Instruction Manual Model GX 82 Portable Three Gas Monitor Part Number 71 0000RK Edition Second Released November 1997...

Page 2: ...eplacement costs local repair costs transportation costs or contingent expenses incurred without our prior approval THIS WARRANTY IS IN LIEU OF ANY OTHER WARRANTIES AND REPRESENTATIONS EXPRESSED OR IM...

Page 3: ...Start up 5 Normal Operation 6 Alarms 7 Safety Self Check 8 Emergency Operation 8 Calibration 9 Supplies and Equipment 9 Calibrating the GX 82 9 Calibration with the RKI Four Gas Cylinder 11 Maintenanc...

Page 4: ...iv Model GX 82 Operator s Manual...

Page 5: ...st follow the instructions and warnings in this manual to assure proper and safe operation of the GX 82 Description The RKI Model GX 82 is an advanced portable instrument for simultaneous detection of...

Page 6: ...ntered stainless steel flame arrestor that permits the atmosphere to diffuse inward but prevents flame to pass outward when an explosive atmosphere is encountered The two elements form half of a balan...

Page 7: ...nsors are exposed to a continuous bias current even when the power is off Control Panel The control panel is at the top of the GX 82 for easy access when you carry the GX 82 The control panel contains...

Page 8: ...batteries A plastic strap with a snap fastener is included to secure the cover in place for everyday use The GX 82 uses two C size cells Standard alkaline batteries will run the GX 82 for approximate...

Page 9: ...operation 2 Verify that the GX 82 is operating correctly To easily verify correct operation of the GX 82 exhale over the diffusion grill When the oxygen reading drops below the decreasing oxygen alarm...

Page 10: ...sounds The GX 82 displays D XXX ppm XXXM This display indicates the H2S or CO average exposure D XXX ppm and duration of the measurement in minutes XXXM The GX 82 returns to normal operation after fi...

Page 11: ...he detection range of the oxygen sensor 40 0 a steady alarm sounds and the oxygen reading is replaced by the word OVER H2S For GX 82 s with H2S detection if H2S exceeds the first alarm setting 10 0 pp...

Page 12: ...ctive sensors The GX 82 performs normal monitoring functions for the active sensors for 13 minutes when it performs a self diagnosis then shows a malfunction alarm To continue using the GX 82 turn off...

Page 13: ...ontrol increases the reading Ignore alarms during the calibration procedure Calibrating the combustibles channel 1 Take the GX 82 to a non hazardous environment before calibrating 2 Turn on the GX 82...

Page 14: ...he oxygen sensor to a know oxygen free sample such as 100 nitrogen available from RKI 2 Watch the display to verify that the O2 reading decrease to 0 0 If the reading does not reach 0 0 adjust the O2...

Page 15: ...it to warm up for five minutes 3 Adjust the Auto Zero Span as described in Normal Operation on page 6 4 Use the supplied tubing to connect the calibration cup to the regulator 5 Attach the calibration...

Page 16: ...circuits to show a normal response CO Filter CO sensors are equipped with an activated carbon filter that removes H2S and most hydrocarbons to limit interference with the CO measurement To install th...

Page 17: ...H2S or CO sensors have been replaced to show a normal response then calibrate Sensor Replacement 1 Place the GX 82 in a non hazardous area 2 Verify that the GX 82 is turned off 3 Press in the cover la...

Page 18: ...clips 7 Plug the sensor block into the end of the extender cable then close the spring clips 8 Slide the sensor cover over the block on the end of the extender cable then slip on the rubber guard 9 S...

Page 19: ...e draw pump allows you to continuously draw and monitor a remote area The sample draw pump includes hose probe pump powered by two AA size batteries NOTE The sample draw pump s batteries have a life s...

Page 20: ...r Description 21 1070RK Battery cover with retaining strap 33 7101RK Filter charcoal 33 7104RK Filter retainer 61 0220RK Combustible gas sensor 65 1051RK Oxygen sensor 65 2003RK Carbon monoxide sensor...

Page 21: ...Appendix B RKI Model GX 82 Relative Response Chart NOTE The following response chart is based on a GX 82 that is calibrated to methane 0 10 20 30 40 50 60 70 80 90 100 0 10 20 30 40 50 60 70 80 90 10...

Page 22: ...18 Appendix B RKI Model GX 82 Relative Response Chart Model GX 82 Instruction Manual...

Reviews: