Preventive Maintenance • 53
5. Remove the fuse holder from the socket, then remove the fuse from the
fuse holder.
CAUTION: Verify that the replacement fuse is the same type and rating as
the fuse you are replacing.
6. Install the appropriate replacement fuse in the fuse holder, then place
the fuse holder in the socket.
7. Push the fuse holder into the socket, then turn the holder 1/4 turn
clockwise to secure it in the socket.
8. Turn on or connect mains power to the Beacon 200.
9. Place the Beacon 200’s power switch in the ON position, then verify
that the PILOT light is on.
10. Close and secure the housing door.
Preventive Maintenance
Preventive maintenance of the Beacon 200 consists of daily, monthly, and
quarterly procedures to ensure that the detector head is operating properly.
See the detector head instruction manual(s) for preventive maintenance
procedures that apply to the detector head(s) supplied with your Beacon
200.