65-2434RK CO Transmitter • 13
Returning to Normal Operation
1.
Remove the voltmeter leads from the amplifier test points.
2.
Secure the cover to the enclosure.
3.
When the display reading falls below the alarm setpoints, return the controller to
normal operation.
4.
Verify that the controller display reading decreases and stabilizes at 0 ppm CO.
5.
Store the components of the calibration kit in a safe and convenient place.
Parts List
Table 6 lists replacement parts and accessories for the CO channel.
Table 3: Parts List
Part Number
Description
06-1248RK
Sample tubing (order by the foot)
07-0203RK
Retaining boot (for filter)
33-7101RK
Filter (charcoal)
57-0035RK-01
Amplifier PCB, CO
65-2434RK
CO transmitter, complete
71-0061RK
65-2434RK CO Transmitter Manual (this document)
81-0064RK
Calibration cylinder (50 PPM CO in air; 34 liter)
81-0076RK-01
Zero air calibration cylinder (34 liter)
81-1050RK
Regulator with gauge and knob, 0.5 LPM, for 17 liter and 34 liter
steel calibration cylinders
81-1109RK
Calibration cup
ES-1531-CO
CO replacement sensor
www.
.com
1.800.561.8187