background image

 

65-2424RK H

2

S Transmitter • 11

CAUTION:

Allow the transmitter to warm up for 5 minutes before you continue with the next 

section, “Setting the Zero Signal”.

Setting the Zero Signal

NOTE:

If you can verify that the detector is in a fresh air environment (environment known to 
be of normal oxygen content and free of toxic and combustible gases), it is not 
necessary to apply zero air when verifying or setting the fresh air reading.

The procedure below describes applying zero emission air, usually called zero air, using a 
calibration kit that includes a calibration cup, calibration gas, sample tubing, and a fixed flow 
regulator with an on/off knob. RKI Instruments, Inc. recommends using a 0.5 LPM (liters per 
minute) fixed flow regulator.

1. Remove the enclosure’s cover by unscrewing the four cover screws, one at each corner.

2. Set a voltmeter to measure in the millivolt (mV) range.
3. Plug the voltmeter leads into the test points on the amplifier PCB. Plug the positive lead 

into the test point labeled 

CAL+

; plug the negative lead into the test point labeled 

CAL-

.

4. Screw the regulator into the zero air calibration cylinder.
5. Use the sample tubing to connect the regulator to the calibration cup.

6. Turn the regulator’s on/off knob counterclockwise to open it.

7. Hold the calibration cup against the H

2

S sensor’s face. Leave the rubber boot and sealing 

spacer on the sensor when calibrating. Hold the cup on gently to allow gas to escape and 
avoid pressurizing the sensor’s sensing face.

8. Allow the gas to flow for 2 minutes.

9. Verify a voltmeter reading of 100 mV (±2 mV).
10. If necessary, use a flat-blade screwdriver to adjust the zero pot until the voltmeter reading 

is 100 mV (±2 mV).

11. Turn the regulator on/off knob clockwise to close it.
12. Remove the calibration cup from the sensor face.

13. Unscrew the regulator from the zero air calibration cylinder. For convenience, leave the 

sample tubing connected to the regulator and the calibration cup.

14. Store the components of the calibration kit in a safe and convenient place.

15. Remove the voltmeter leads from the test points.

16. Secure the cover to the enclosure.

Summary of Contents for 65-2424RK

Page 1: ...www rkiinstruments com 65 2424RK Hydrogen Sulfide Transmitter Operator s Manual Part Number 71 0062RK Revision B Released 6 12 20...

Page 2: ...alibration and maintenance of the detector is essential for proper operation and correct readings Please calibrate and maintain this detector regularly Frequency of calibration depends upon the type o...

Page 3: ...ent costs local repair costs transportation costs or contingent expenses incurred without our prior approval THIS WARRANTY IS EXPRESSLY IN LIEU OF ANY AND ALL OTHER WARRANTIES AND REPRESENTATIONS EXPR...

Page 4: ...ring the H2S Transmitter to a Controller 9 Startup 10 Introducing Incoming Power 10 Setting the Zero Signal 11 Maintenance 12 Preventive Maintenance 12 Troubleshooting 12 Replacing Components of the H...

Page 5: ...ransmitter WARNING When using the 65 2424RK you must follow the instructions and warnings in this manual to assure proper and safe operation of the 65 2424RK and to minimize the risk of personal injur...

Page 6: ...trical output that is proportional to the detection range of the transmitter Sealing Spacer The ring shaped plastic sealing spacer fills a gap in the rubber boot that is installed on the sensor The ru...

Page 7: ...the span pot Use the zero pot to adjust the transmitter s target gas free output during the start up and calibration procedures CAUTION There is a third potentiometer on the PCB the null potentiomete...

Page 8: ...m to perform start up maintenance and calibration procedures Select a site where the target gas is likely to be found first NOTE If your application does not require a specific mounting site mount the...

Page 9: ...to the interconnect terminal strip as follows see Figure 3 Connect the positive wire to the terminal labeled V Connect the feedback wire to the terminal labeled FB CAUTION Leave the shield drain wire...

Page 10: ...p This section describes procedures to start up the H2S transmitter and place the transmitter into normal operation Introducing Incoming Power 1 Complete the installation procedures described earlier...

Page 11: ...ositive lead into the test point labeled CAL plug the negative lead into the test point labeled CAL 4 Screw the regulator into the zero air calibration cylinder 5 Use the sample tubing to connect the...

Page 12: ...counter with the H2S transmitter NOTE This troubleshooting guide describes transmitter problems only See the controller operator s manual for problems you may encounter with the controller Fail Condit...

Page 13: ...cludes procedures to replace the H2S sensor amplifier PCB and charcoal filter Replacing the Sensor 1 Turn off the controller 2 Turn off or unplug incoming power at the power source end 3 Remove the en...

Page 14: ...emoved in step 5 to secure the PCB to the enclosure 10 Reconnect the wiring from the controller to the interconnect terminal strip as shown in Table 2 and Figure 3 11 Carefully plug the H2S sensor wit...

Page 15: ...ing set the response reading and return to normal operation It describes the test using a calibration kit that includes a calibration cup calibration gas sample tubing and a fixed flow regulator with...

Page 16: ...e Reading 1 Verify that the calibration gas is representative of the transmitter s target gas 2 For toxic gas cylinders like H2S it is important to vent the regulator while installing it onto the cyli...

Page 17: ...on 06 1248RK Sample tubing order by the foot 07 0203RK Retaining boot for filter 14 2101RK Sealing spacer 57 0035RK 02 Amplifier PCB H2S 65 2424RK H2S transmitter complete 71 0062RK 65 2424RK H2S Tran...

Reviews: