![RKI Instruments 65-2380 Operator'S Manual Download Page 18](http://html1.mh-extra.com/html/rki-instruments/65-2380/65-2380_operators-manual_1466807018.webp)
18
•
Calibration Frequency
NOTE:
When a transmitter is first powered up with a new amplifier, the initial
output may be either high or below zero depending on the setting of the zero
pot. Be sure to make arrangements so that this does not cause unwanted
alarms.
10. Turn on power to the controller.
11. Turn on the controller and place it into normal operation.
12. Allow the transmitter to warm-up for 5 minutes.
13. Calibrate the transmitter as described in “Calibration” on page 18 of this manual.
Calibration Frequency
Although there is no particular calibration frequency that is correct for all applications, a
calibration frequency of every 6 months is adequate for most MPS combustible gas transmitter
applications. Unless experience in a particular application dictates otherwise, RKI Instruments,
Inc. recommends a calibration frequency of every 6 months.
If an application is not very demanding, for example detection in a clean, temperature controlled
environment where calibration adjustments are minimal at calibration, then a calibration
frequency of every 9 to 12 months is adequate.
If an application is very demanding, for example if the environment is not well controlled, then
more frequent calibration than every 6 months may be necessary.
Calibration
This section describes how to calibrate the combustible gas transmitter. It includes procedures to
prepare for calibration, set the zero reading, set the response reading, and return to normal
operation.
WARNING: The controller is not an active gas monitoring device during the calibration
procedure.
NOTE:
The following procedure assumes the use of a calibration kit which includes a
calibration gas cylinder, a 0.5 LPM fixed flow regulator with an on/off knob, a
calibration cup for the detector, and a short piece of sample tubing to connect the
regulator to the calibration cup.
NOTE: During calibration, the detector must be calibrated with a cylinder of methane in a
background of real air containing argon and CO
2
. Only use the RKI-provided
calibration cylinder 81-9070RK-03.