65-2340RK Toxic Gas Transmitter • 13
Start Up
This section describes procedures to start up the toxic gas transmitter and place the transmitter
into normal operation.
Introducing Incoming Power
1. Complete the installation procedures described earlier in this manual.
2. Verify that the power wiring to the controller is correct and secure. Refer to the controller
operator’s manual.
3. Turn on power to the controller.
4. Turn on the controller.
5. Verify that the controller is on and operating properly. Refer to the controller operator’s
manual.
CAUTION:
Allow the transmitter to warm up for 5 minutes before you continue with the next
section, “Setting the Zero Signal”.
Setting the Zero Signal
CAUTION:
If you suspect the presence of toxic gas in the monitoring environment, use the
calibration kit and the zero air calibration cylinder to introduce “fresh air” to the
detector and verify an accurate zero setting. See “Calibration” on page 18 for
instructions to apply zero air when setting the zero signal.
1. Verify that the transmitter is in a fresh air environment (environment known to be free of
toxic gas).
2. Unscrew and remove the junction box cover from the junction box.
3. Set a voltmeter to measure in the millivolt (mV) range.
4. Plug the voltmeter leads into the test points on the amplifier. Plug the positive lead into the
red + test point; plug the negative lead into the black
-
test point.
5. Verify a voltmeter reading of 100 mV (±2 mV).
6. If necessary, use a small flat-blade screwdriver to adjust the zero pot until the voltmeter
reading is 100 mV (±2 mV).
7. Remove the voltmeter leads from the test points.
8. Secure the junction box cover to the junction box.