6 • 35-3001A-14 Methane Sample-Draw Detector
IR CH
4
Amplifier
Each IR CH
4
amplifier converts the electrical output from the IR CH
4
sensor to a 4 to 20 mA signal that
corresponds to the detection range and transmits the signal to a gas monitoring controller. A label on the
amplifier indicates the detector drive current. This drive current is factory set. One amplifier is mounted on
the top middle edge of the main circuit board. The other is mounted in the bottom left corner of the main
circuit board. Each amplifier consists of the zero pot, span pot, controller terminal strip, detector terminal
strip, and test points.
Figure 3: IR CH
4
Amplifier Component Location
Zero Pot
The zero pot is located in the upper right corner of the amplifier (see Figure 3). Use a small flat blade
screwdriver to turn the zero pot’s adjustment screw and adjust the amplifier’s zero (fresh air) output during
the start-up and calibration procedures. Turn the adjustment screw clockwise to increase the fresh air output
and counterclockwise to decrease the fresh air output.
Span Pot
The span pot is located below the zero pot (see Figure 3). Use a small flat blade screwdriver to turn the span
pot’s adjustment screw and adjust the amplifier’s gas response output during the calibration procedure. Turn
the adjustment screw clockwise to increase the span output and counterclockwise to decrease the span
output.
CAUTION:
The amplifier includes additional pots. They are factory-set. Do not adjust them.
Controller and Detector Terminal Strips
The controller terminal strip and detector terminal strip are three- and four-position plug-in style terminal
strips, respectively. The controller terminal strip is located on the right side of the amplifier and the detector
terminal strip is to the left of it. Both terminal strips are factory wired to the sensor and main circuit board.
Test Points
The test points are on the left side of the amplifier (see Figure 3). The test points produce a 100 mV to 500
mV output that corresponds to the sample-draw detector’s 4 to 20 mA output. Use the test points and a
voltmeter to measure the amplifier’s output during the start-up and calibration procedures. The black test
point in the upper left corner is the negative (-) test point and the red test point to the left of the zero and span
pots is the positive (+) test point.
Factory Set Pots
Sensor
Current
148mA
W
LEL
G
B
R
Test Points
100-500 mV Range
Zero Pot
Detector
Terminal Strip
Span Pot
Controller
Terminal Strip
S
PWR / SIG