35-3001A-01H Hydrogen Sample-Draw Detector • 7
Installation
This section describes procedures to mount the sample-draw detector in the monitoring
environment and wire the sample-draw detector to a controller.
Mounting the Sample-Draw Hydrogen Detector
1. Select the mounting site. Consider the following when you select the mounting site:
•
Is there enough room to open the housing door and make wiring and tubing connections at
the bottom of the housing?
•
Make sure there is sufficient room to perform start-up, maintenance, and calibration
procedures.
•
Are the flowmeter and status LEDs visible?
Figure 4: Outline and Mounting Dimensions
Ø .30 x .50 (4X) M OUNTING
.78
1.13
.40
Pilot Fail
.85
.80
4.50
NOTE: Housing is 4.3 inches d eep
8.88
4.00
6.50
35-3001
www.rkiinstruments.com
Sample Drawing
Gas Detector
8.50