22 • 35-3001-13-02 Carbon Dioxide/Oxygen Sample-Draw Detector
4. Follow the directions in the controller’s operator’s manual for setting the span.
5. Disconnect the sample tubing from the inlet line.
6. Unscrew the regulator from the carbon dioxide calibration cylinder.
Returning to Normal Operation
1. Wait approximately one minute to allow the carbon dioxide reading to stabilize.
2. Follow the instructions in the controller’s operator’s manual to exit the calibration mode.
3. Store the components of the calibration kit in a safe and convenient place.
Calibration, Oxygen Sensor
This section describes how to calibrate the oxygen sensor. It includes procedures to prepare for calibration, set
the fresh air reading, set the zero reading, and return to normal operation.
NOTE:
Calibrating the sample draw detector may cause alarms. Be sure to put the controller into its calibration
program or disable external alarms before continuing.
NOTE:
This procedure describes calibration using a demand flow regulator.
Preparing for Calibration
NOTE:
If you can verify a fresh air environment, it is not necessary to use a zero air calibrating sample to set
the fresh air reading at the controller.
1. Follow the instructions in the controller’s operator’s manual for entering calibration mode.
2. Screw the regulator into the zero air calibration cylinder.
Setting the Fresh Air Reading
1. Connect the sample tubing from the demand flow regulator to the sample-draw detector’s inlet fitting.
This
step is not necessary if you verified a fresh air environment earlier in this procedure.
2. Allow the sample draw detector to draw sample for one minute.
3. Follow the directions in the controller’s operator’s manual for setting the fresh air reading.
4. Disconnect the sample tubing from the inlet fitting.
5. Unscrew the regulator from the zero air calibration cylinder.
Setting the Zero Reading
1. Screw the demand flow regulator onto the 100% N
2
calibration cylinder.
2. Connect the sample tubing from the regulator to the inlet line.
3. Allow the sample-draw detector to draw the calibrating sample for 1 minute.
4. Follow the directions in the controller’s operator’s manual for setting the zero (oxygen free) reading.
5. Disconnect the sample tubing from the sample-draw detector’s inlet fitting.
6. Reconnect the incoming sample line to the inlet fitting.
7. Unscrew the regulator from the calibration cylinder.