16 • 35-3001-06H-DIL Hydrogen/Oxygen Sample-Draw Detector
Replacing Components of the Sample-Draw Detector
This section includes procedures to replace the hydrogen sensor, oxygen sensor, hydrophobic filter,
and particle filter.
Replacing the Hydrogen Sensor
1. Turn off the controller.
2. Turn off power to the controller.
3. Open the housing door of the sample-draw detector.
4. Unscrew and remove the two screws that secure the sensor retaining plate, then lift the plate,
connector, and sensor out of the housing.
5. Unplug the connector from the sensor.
6. Verify that you are using the correct replacement sensor (NC-6244A is printed on the sensor),
then plug the sensor into the connector.
7. Place the sensor in the hydrogen sensor cavity, then position the retaining plate on the two
standoffs.
8. Secure the retaining plate to the standoffs with the two screws you removed in step 4.
9. Close and latch the housing door.
10. Turn on power to the controller.
11. Turn on the controller.
CAUTION:
Allow the replacement sensor to warm up for 5 minutes before you continue.
12. Calibrate the replacement sensor as described in “Calibration, LEL Sensor” on page 19.
Replacing the Oxygen Sensor
1. Turn off the controller.
2. Turn off power to the controller.
3. Open the housing door of the sample-draw detector.
4. Loosen the two screws on the oxygen terminal strip and remove the lugs and wires from the
terminal strip.
5. Unscrew and remove the two screws that secure the sensor retaining plate, then lift the plate
and sensor out of the flow block.
There are two gaskets at the bottom of the flow block. Be sure the gaskets stay in place.
6. Place the replacement sensor in the oxygen flow block, then position the retaining plate on the
two standoffs.
7. Secure the retaining plate to the standoffs with the two screws you removed in step 4.
8. Guide the wires and lugs of the replacement sensor to the oxygen terminal strip and insert each
lug into the appropriate terminal. See Figure 4 on page 11 for wiring connections. Tighten the
screws on the terminal strip.
9. Close and latch the housing door.
10. Turn on power to the controller.
11. Turn on the controller.
CAUTION:
Allow the replacement sensor to warm up for 5 minutes before you continue.