4 • 35-3001-01H-DIL Hydrogen Sample-Draw Detector
External Components
This section describes the sample-draw detector’s external components.
Housing
The sample-draw detector’s fiberglass housing is weather- and corrosion-resistant. It is suitable for
installation where general purpose equipment is in use.
The housing door is hinged on the left side and is secured by two latches on the right side. The
flowmeter and status LEDs are visible through a window in the housing door.
Four mounting feet are attached to the back of the housing (one at each corner). Use the mounting
feet to install the housing to a vertical surface.
Sample Fittings
The sample fittings are located on the left side of the bottom of the housing. The inlet fitting is near
the front of the housing and the exhaust fitting is near the back of the housing. The sample fittings
accept 1/4 in. rigid tubing. See the Installation section on page 7 to connect tubing to the sample
fittings.
Particle Filter
A particle filter with a tubing stub on one end is shipped with the instrument but it is not factory
installed. If the particle filter is installed directly to the inlet fitting, the tubing stub must be used. If
the particle filter is installed somewhere else, like at the end of the inlet line, the tubing stub can be
used or removed.
Conduit Hub
One 3/4” conduit hub is located on the right side of the bottom of the housing. It is used for routing
wiring into the housing by using conduit or an appropriate cable bushing.
Internal Components
This section describes the sample-draw detector’s internal components (see Figure 1). Figure 2
illustrates how the gas sample moves through the flow system.
Figure 2: Hydrogen Sample-Draw Detector Flow Diagram
LEL
Sensor
Flowmeter
Hydrophobic Filter
Pump
Exhaust
Pressure Switch
Flowmeter PCB
Sensor Flow
Control Valve
Restrictor
Inlet
Dilution Fitting,
approx. 1:1 ratio
Particle Filter