SR-496 Surface Resistance Meter 

Instruction Manual 

 

The surface resistance can be measured by attaching two parallel electrodes of the instrument to the surface 
of the measured object. Thus you can determine whether the measured object is a conductor, anti-static 
material, or an insulator. The instrument is particularly suitable for a variety of anti-static areas. 

 

A.

 

Specifications of SR-496 

Resistance Measurement Range 

0.1 x 10

~ 1.5 x 10

12 

Ω 

Resistance Measurement Accuracy 

<±10% 

Response Time 

Approx. 1 sec. 

Temperature Measurement Range 

0

℃ 

~ +50

℃ 

(+32

℉ 

~ +122

℉)

 

Temperature Measurement Accuracy 

±1

  (±1.8

℉)

 

Power Supply 

One 9V battery 

Dimension 

64 x 35 x 126mm (2.5” x 1.2” x 5.0”) 

Weight 

Approx. 156g (battery included) 

 

B.

 

Material Resistance Range 

Conductor 

<10

6

 

Anti-Static Material 

10

~ 10

11

 

Insulator 

10

12

 

The signal is weak when measure surface resistance of 10

11

 Ω or more, in order to ensure stable and 

reliable measurement, it is necessary to make environment where electromagnetic interference is small. 
During the measurement process, make sure that no person walks within 3 meters. 

 

C.

 

Function 

  Surface resistance test   

  Three ways to display the surface resistance or material properties simultaneously 

  Environment temperature measurement   

  Temperature unit conversion   

  Data hold   

  LCD backlight   

  Auto turn off 

 

D.

 

LCD Display

 

(Please see the right photo.)

 

 

Insulation material indicator 

 

Anti-static material indicator 

 

Conductor indicator 

 

Resistance value 

 

Temperature value 

 

Resistance measurement indicator 

 

Data hold indicator 

 

Battery indicator 

 