background image

Page 6

DLS360 User Guide

To output a 2kHz tone simply turn the device on, select your desired wavelength, then press the “2kHz” button. The 

wavelength indicator will begin blinking to confirm to the user that 2kHz tone is being output. Plugging in a power meter 

like the RP450 will display an onscreen notification and output a loud beep to alert the user at the other end that a tone 

has been detected. To disable the tone simply press the 2kHz button again.

Using 2kHz

Maintenance

The DLS360 utilizes a physical fiber connection at the output port. This ensures a steady power level for performing 

insertion loss tests.

Be aware that any test jumpers must be inspected and cleaned before plugging into the DLS360 unit. If soiled or damaged 

connectors are inserted, they can cause damage to the DLS360 output port and the unit may need to be repaired. 

The test jumper has 

physical contact with the 

ferrule inside the DLS360.

To clean the ferrule  

inside the DLS360:

•  Unplug the test jumper

•  Unscrew the adapter until it spins 

freely, then pull

•  Inspect the ferrule with an 

approved microscope, clean if 

necessary, and replace adapter

Caring for the DLS360 Output Port

Singlemode fibers usually have either blue (UPC) or green (APC) connectors. The DLS360 Laser source uses a UPC 

connector so neither beige nor green connectors are appropriate.

DO NOT plug an APC connector into the DLS360 unit. This can cause irreparable damage to 

both connectors.

Summary of Contents for Miller DLS360

Page 1: ...Page 1 DLS360 User Guide DLS360 Dual Laser Source User Guide...

Page 2: ...Page 2 DLS360 User Guide Contents Introduction 3 Device Overview 3 Important Safety Information 4 Basic Testing 5 Using 2kHz 6 Maintenance 6 Certifications and Contact Information 7...

Page 3: ...the DLS360 ON and OFF The unit will turn off automatically after 15 minutes by default To disable this timeout hold the power button for at least 5 seconds when turning ON 2kHz Button Toggles 2kHz ton...

Page 4: ...d non reflective surfaces to prevent reflection of laser Electric Shock Hazard Pay attention to proper battery polarity Do not mix battery types or manufacturers Do not open the unit with the exceptio...

Page 5: ...pport see the final page of this document for contact information Transmitting Light The DLS360 transmits either the 1490nm or 1625nm wavelength on Multimode fiber Be sure to use a test jumper to mate...

Page 6: ...s Be aware that any test jumpers must be inspected and cleaned before plugging into the DLS360 unit If soiled or damaged connectors are inserted they can cause damage to the DLS360 output port and the...

Page 7: ...ee recalibration within 2 years of the date of purchase To start the calibration process simply call or email technical support Calibration Certificates The DLS360 comes with a 2 year warranty for any...