TECHNICAL DATA 

(

According to ANSI standard S3.22 1996

)

Maximum-OSPL90

130 dB (1200 Hz)

HFA-OSPL90

127 dB

HFA-full-on Acoustic Gain

52 dB

Reference Test Gain

50 dB

Frequency Range

100 Hz to 6800 Hz

Equivalent Input Noise Level

20 dB

Total Harmonic Distortion

500 Hz: 8.0% 800 Hz: 3.0% 1600 Hz: 3.0%

HFA-SPLIV

112 dB at 31.6 mA/m

AGC

OPC (Output AGC)

Attack/Recovery Time

Attack: 0.25 msec/Recovery: 32 msec

Operating Switch

O-T-M

Output Limiting Control

OPC (range: 24 dB)

Tone Control

TONE H, TONE L

Gain Control

Main VR (range: 38 dB)

Battery Type/Supply Voltage

13 (13P)/1.3 V

Battery Current/Battery Life

13:

1.24 mA/Approx. 210 h

13P: 1.24 mA/Approx. 160 h

Dimensions/Weight

3.75 × 1.27 × 0.88 cm/4.2 g

 

(excluding battery)

(Typical value with E1 hook)

TECHNICAL DATA 

(

)

Reference Test Frequency

1600 Hz

OSPL90

131 dB

500 Hz

126 dB

Peak

136 dB

Full-on Acoustic Gain

55 dB

Equivalent Input Noise Level

17 dB

Total Harmonic Distortion

500 Hz: 2.0% 800 Hz: 1.0% 1600 Hz: 1.0%

Induction Coil Sensitivity

89 dB at 1 mA/m

AGC

OPC (Output AGC)

Attack/Recovery Time

Attack: 0.25 msec/Recovery: 32 msec

Operating Switch

O-T-M

Output Limiting Control

OPC (range: 24 dB)

Tone Control

TONE H, TONE L

Gain Control

Main VR (range: 38 dB)

Battery Type/Supply Voltage

PR48 (PR48P)/1.3 V

Battery Current / Battery Life

PR48:

1.24 mA/Approx. 210 h

PR48P: 1.24 mA/Approx. 160 h

Dimensions/Weight

3.75 × 1.27 × 0.88 cm/4.2 g

 

(excluding battery)

(Typical value with E1 hook)

According to IEC standard Pub. 60118-0-1983
Amendment 1-1994

Opening / closing the control cover

1. Insert a screwdriver into the

notch of the control cover, as
shown in Fig. 1, and push the
cover down.
(Only the top section of the
cover will open.)

2. When the top of the cover opens, insert the screwdriver into the

clearance, as shown in Fig. 2, and push the cover out to open it fully.

3. To close the cover, push the entire cover in with finger.

Adjusting the controls

The scales of the tone control (TONE H, TONE L)
and output limiting control (OPC) are shown in the
illustration (Values shown in bold are printed on
the unit.). Use the numbers given on a scale as a guide when adjusting
the control. For information on volume control positions, see the section
"Volume control positions for 2000 Hz hearing level".

z

Tone control (TONE H)

As you turn the control from "N" towards "H",
the low range is attenuated and the high range
is emphasized. To make the adjustment using
numerical values, calculate the value as
follows.

TONE H setting = HL (2000 Hz) - HL (500 Hz)

z

Tone control (TONE L)

As you turn the control from "N" towards "L", the high range is
attenuated and the low range is emphasized. Each increment of the

printed scale corresponds to a change of about
4 dB at 4000 Hz.
Normally, the control should be left in the factory-
default position (N). If acoustic feedback (howling)
is a problem or when the hearing level in the low
range seems notably worse than in the high range, turn the control
towards "L".

z

Output limiting control (OPC)

Turning the control towards the left reduces the
maximum output level of the hearing aid. Set the
control to a position which matches the average
hearing level. The average hearing level can be
determined according to the following equation.

Average hearing level = 

HL (500 Hz) + { 2 × HL (1000 Hz) } + HL (2000 Hz)

4

* About batteries

The generally recommended battery type for the
HB-41 is the PR48, but when the OPC control is
set in the range indicated by the arrows in blue,
the PR48P should be used. With the PR48, the
sound may briefly be interrupted at high levels. The PR48P eliminates
this problem, but this battery type has a slightly shorter service life.

Volume control positions for 2000 Hz hearing level