7 Maintenance
7-1. Maintenance intervals and items
- 32 -
Main Services
Power Supply
Check
: Checks the power supply voltage.
Verifies that the power lamp lights up.
(Verifies that relevant points can be identified on the system.)
(When a UPS (uninterruptible power system) is used, checks the operation with the UPS.)
Concentration
Display Check
: Verifies that the concentration display value is zero (or 20.9 vol% on the oxygen deficiency
meter) by using the zero gas.
Performs the zero adjustment (fresh air adjustment) if the reading is incorrect.
Flow Rate
Check
: Checks the flow rate indicator to find abnormalities.
Checks the flow rate by using an external flow meter to verify the correctness of the flow
rate indicator on the gas detector. If the flow rate is incorrect, performs the flow rate
adjustment.
Filter Check
: Checks the dust filter for dust or clogging.
Replaces a dirty or clogged dust filter.
Alarm Test
: Inspects the alarm circuit by using the alarm test function.
Checks the alarm lamps. (Checks ALM activation.)
Checks the external alarm. (Checks the activation of the external alarm, such as a
buzzer.)
Span
Adjustment
: Performs the span adjustment by using the calibration gas.
Gas Alarm
Check
: Checks the gas alarm by using the calibration gas.
Checks the alarm. (Checks triggering of alarm when the alarm setpoint is reached.)
Checks the delay time. (Checks time to delay until the alarm is triggered.)
Checks the alarm lamps. (Checks ALM activation.)
Checks the external alarm. (Checks the activation of external alarms, such as a
buzzer and reset signal.)
Cleaning and
Repair of
Device
(Visual
Diagnosis)
: Checks dust or damage on surface, cover or internal parts of the detector, cleans and
repairs such parts of the detector.
Replaces parts which are cracked or damaged.
Device
Operation
Check
: Uses the keys to check the operation of functions and parameters.
Replacement
of Consumable
Parts
: Replaces consumable parts, such as a sensor, filter and pump.
Summary of Contents for SD-1DOX
Page 1: ...PT2E 3001 Smart Transmitter Gas Detector Head SD 1DOX TYPE MED EX Operating Manual PT2E 179...
Page 5: ......
Page 6: ......
Page 7: ......
Page 8: ......
Page 9: ......
Page 10: ......
Page 12: ......
Page 13: ......
Page 14: ......
Page 15: ......
Page 16: ......
Page 17: ......
Page 18: ......
Page 19: ......
Page 20: ......
Page 21: ......
Page 22: ......
Page 23: ......
Page 24: ......
Page 25: ......
Page 26: ......
Page 27: ......
Page 28: ......
Page 31: ......
Page 32: ......
Page 35: ......
Page 36: ......
Page 37: ......
Page 38: ......
Page 39: ......
Page 40: ......
Page 41: ......
Page 43: ......
Page 44: ......
Page 45: ......
Page 47: ......
Page 51: ......
Page 53: ......
Page 56: ......