Operating Precautions
This detector is an oxygen deficiency detector that detects oxygen in the air and triggers an oxygen
deficiency alarm.
The oxygen deficiency detector is a safety unit, not an analyzer or densitometer which performs
quantitative/qualitative analysis/measurement for oxygen.
Please fully understand the following points before using it, so that it can be used properly.
1. The readings of the detector fluctuate slightly in response to changes in the air pressure.
In particular, be careful of alarm activation when a low air pressure is brought in by typhoon. In addition,
it may be fluctuated by environmental (temperature, humidity etc.) changes in the installation site.
2. The alarm must be set within a range where the performance of the detector can be ensured.
The standard alarm setpoint value is 18 vol%.
3. If use in safety unit, the analog signal output of the detector must be used for an indicator or external
recorder.
If these outputs are used to control other units, we shall not be responsible for any malfunctions.
4. For maintenance of the detector, it must go through a regular maintenance, including replacement and
adjustment of the regular replacement parts as specified in the operating manual. In addition, because
this is a safety unit, it is recommended that a regular maintenance and span adjustment are performed
every six months.
Summary of Contents for SD-1DOX-AS
Page 5: ......
Page 6: ......
Page 7: ......
Page 8: ......
Page 9: ......
Page 10: ......
Page 12: ......
Page 13: ......
Page 15: ......
Page 16: ......
Page 17: ......
Page 18: ......
Page 20: ......
Page 21: ......
Page 22: ......
Page 23: ......
Page 24: ......
Page 25: ......
Page 26: ......
Page 27: ......
Page 28: ......
Page 29: ......
Page 30: ......
Page 31: ......
Page 32: ......
Page 33: ......
Page 34: ......
Page 37: ......
Page 38: ......
Page 41: ......
Page 42: ......
Page 43: ......
Page 44: ......
Page 45: ......
Page 46: ......
Page 47: ......
Page 48: ......
Page 49: ......
Page 50: ......
Page 51: ......
Page 52: ......
Page 53: ......
Page 54: ......
Page 61: ...10 Product Specifications 10 1 List of specifications 59 Outline Drawings Adapter A NPT1 2...
Page 65: ...10 Product Specifications 10 1 List of specifications 63 Outline Drawings Cable gland G3 4...
Page 66: ......
Page 69: ......