9
3−4.
Check of the flowmonitor
When turn on power, the pump starts working. Check that it works properly by the flowmonitor
(float) of filter tube with the flowmonitor.
When sucking properly (Can see the float)
Not sucking properly (Cannot see the float)
3−5.
Zero adjustment
Make zero adjustment by removing the gas sampling tube. After the detection starts, see the
flowmonitor and check that the air is sucked properly. Then, check that the reading gets stable
and make zero adjustment by sucking the fresh air into instrument.
The operation procedure is shown below.
0.0%LEL 20.9%
(
Gas display
)
↓
HOLD ZERO KEY
Hold pressing
ZERO
for approx. 3 sec.
↓
APPLY FRESH AIR
After checking that fresh air is sucked into instrument
↓
and then, press
ENTER
switch.
ADJUSTING ZERO
Zero calibration failure
Approx. 5 sec.
↓
①
ZERO SET OK
Approx. 5 sec.
↓
0.0%LEL 20.9%
(
Gas display
)
Float
Float