9
(2) Average value measuring
Average value can be measured by pressing AVG.ST switch while displaying gas
concentration.
・
Three kinds of measuring time are available,
1 minute ・ 3 minutes ・ 15 minutes .
・
The result of measured average value is memorized after power off.
・
Measuring time is a little bit different from the atmosphere temperature.
・
By starting measuring average value newly, the previous data is cleared, and new
measured average value is memorized. If cancel new average value measuring,
the previous data is kept.
・
In case of over 10,000ppm (Indicates "OVER"), the all data is measured
"10,050ppm".
① 700 ppm
(Gas concentration indication.)
Press
AVG.ST
switch.
② 1min:START→ENT
(Blinking measuring time.)
By + - switches, measuring time can be selected.
③ 3min:START→ENT
(Blinking measuring time.)
Press
ENTER
switch, start measuring average value.
④
3min: 800 ppm
(Measuring average value.)
Indicate moment concentration value.
Time count down.
⑤ 58sec: 880 ppm
(Measuring average value.)
Change over from "min" to "sec" when the rest time becomes within 1 minute.
⑥ AVG.: 850 ppm
(Measuring average value end.)
Press
ESC
switch, return to gas concentration indication. (to ①)
Press
AV/PK
switch, change over to peak value indication. (to (4)-③)
Press
AVG.ST
switch, change over again to measuring average value. (to ②)