<ID Display/Selection [ID SELECT]>
Display an ID if it has been registered in advance. Also used to select an ID.
(1) Press the DISPLAY switch and select the ID display/selection from the display/setting mode menu.
The following screens are displayed in turn on the gas monitor.
(2) Press the ENTER switch to set or select an ID.
NOTE
If you do not want to set or select an ID, press the DISPLAY switch to return to the display/setting
mode menu.
On this gas monitor, either of the IDs from ST-ID000 to ST-ID255 has been registered, unless
otherwise specified.
The data logger management program (option) is required to register or change an ID. Please contact
RIKEN KEIKI.
(3) Press either the
▲
or
▼
switch to select an ID.
Every time the
▲
or
▼
switch is pressed, the ID number increases or decreases (000 - 255).
(4) Press the ENTER switch.
(5) When [END] is displayed, the setting is completed.
The display/setting mode menu is displayed again.
(5) After completion, press the DISPLAY switch several times until it returns to the detection mode.