2 Installation
2-1. Requirements
- 11 -
PLU-70
2
Installation
2-1. Requirements
Not only the first-time users but also the users who have already used the product must follow the operating
precautions.
Ignoring the precautions may damage the pyrolyzer unit, resulting in inaccurate gas detection.
<Precautions for Installation Sites>
< Precautions for System Designing >
Using a stable power supply
The pyrolyzer unit must be provided with the following power supply.
Power supply voltage
24 VDC
±
10% (the terminal voltage of the pyrolyzer unit)
Allowed time of
momentary blackout
Up to 10 millisecond
(To recover from the momentary blackout for 10 milliseconds or more, restart the
pyrolyzer unit.)
To ensure continuous operation and activation, install a UPS outside the pyrolyzer unit.
Others
Do not use it with a power supply of large power load or high-frequency noise.
Use a line filter to avoid the noise source if necessary.
CAUTION
•
Install the pyrolyzer unit near the base unit, gas detector head GD-70D series. Since the pyrolyzer
unit is a precision device like the gas detector head, the precautions for installation sites shall conform
to those of the gas detector head. (See the operating manual for the GD-70D series.)
CAUTION
•
After you receive the pyrolyzer unit, start using the pyrolyzer immediately.
CAUTION
•
An unstable power supply and noise may cause malfunctions or false alarms.
•
The designing of a system using the pyrolyzer unit must reflect the contents of the operating manual
of the base unit, gas detector head GD-70D series, and the descriptions in this section.