20
8-3. Optional accessories
Belt clip (with 2 screws)
Extension cable (5m, 10m, 20m, 30m)
Carrying case
[Using the OX-07 with extension cable]
1)
Verify that the OX-07 is off.
2)
Take out the sensor from the instrument body.
3)
Mount the removed sensor to the sensor retainer with the same procedures as sensor
mounting. (See section 2-2.)
4)
Mount the cable retainer to the instrument body as shown below figure, and align the
yellow line with “LOCK” line.
・
When using the OX-07 with extension cable, do not throw sensor part into detection
point, but lower down slowly. Avoid impact to the sensor. Sensor may be damaged.
・
When intending to measure the oxygen content in manhole, inside device, etc by
OX-07 with extension cable, adjust the temperature of the sensor to be the same
temperature as measuring point by leaving it there about 10 minutes. Then, pull it up
to the ground once, turn on the unit again and perform the fresh air adjustment. After
that, lower the sensor down to the detection point and measure oxygen content.
Belt clip
Extension cable
CAUTION
Yellow line
UNLOCK
LOCK
Sensor
Sensor
retainer
Cable retainer