11.0 TYPE-01 SPECIFICATIONS
FOR SOLVENTGASES (vol%)
Selectable measuring gas and range
Measuring gas
(
Display indication
)
Measuring range
(Minimum digit
)
AIR CAL value
Drift amount
Spec. No.
1
Toluene in AIR
(
C7H8 in AIR
)
0
~
2 vol%
(0.002
)
0.000 vol%
0.017 vol%
SPE-1428
2
Methyl ethyl ketone in AIR
(
MEK in AIR
)
0
~
5 vol%
(
0.005
)
0.000 vol%
0.027 vol%
SPE-1429
3
Ethyl acetate in AIR
(
EtAc in AIR
)
0
~
5 vol%
(
0.002
)
0.000 vol%
0.023 vol%
SPE-1430
4
Xylene in AIR
(C8H10 in AIR)
0
~
1 vol%
(0.002)
0.000 vol%
0.014 vol%
SPE-1431
5
Isopropyl alcohol in AIR
(IPA in AIR)
0
~
2 vol%
(0.005)
0.000 vol%
0.033 vol%
SPE-1432
6
Methyl isobutyl ketone in AIR
(MIBK in AIR)
0
~
1 vol%
(0.002)
0.000 vol%
0.016 vol%
SPE-1433
7
Methanol in AIR
(CH3OH in AIR)
0
~
6 vol%
(0.02)
0.00 vol%
0.11 vol%
SPE-1446
8
-
-
-
-
-
Interferometer type : Type 120MM
Indication accuracy : ±( Indication value AIR CAL. value )×0.03 ± Drift amount
AIR CAL value : Concentration indicated on the display when AIR CAL key is pressed.
Drift amount : The amount of indication drifting when temperature change ±10
℃
.
Special caution
1.
This instrument is not explosion proof.
2.
This instrument measures combustible gas around lower explosive limit in this specification. Take enough of
a sample for the measurement.
3.
Measuring gases in this specification are not vaporized so much. The gas concentration may not reach
higher concentration in the measuring range depending on a condition of temperature and pressure.