![Rigol TX1000 RF Application Note Download Page 17](http://html1.mh-extra.com/html/rigol/tx1000-rf/tx1000-rf_application-note_1461070017.webp)
RIGOL
TX1000 RF Demo Kit Application Note
13
Typical Applications
To Test the Measurement Function of the Spectrum
Analyzer
Operation Steps:
1.
Connect the PC and TX1000 using USB cable and connect J10 of TX1000 and the RF input
terminal of the spectrum analyzer using N-SMA cable as shown in Figure 6;
Explanation: when using the DSA800 series spectrum analyzer, you can connect the DSA800
and TX1000 using USB cable without using the PC.
2.
Configure the status of each switch via the control interface.
Switch
PC Control Software Status
DSA800 Control Interface Status
Sw1
Disconnect J1 and connect the
built-in 50 MHz signal
Off
Sw2
Disconnect J3 and J4
Off
Sw3
Disconnect J5 and J6
Off
Sw4
Disconnect J7 and J8
Off
Sw5
Disconnect J12
Off
3.
Set the measurement parameters on the spectrum analyzer to get the measurement results.
Figure 6 Test the Measurement Function of the Instrument