RIGOL
VSA Programming Guide
2-37
:INITiate Command
Command List:
:INITiate:BER:RESTart
Syntax
:INITiate:BER:RESTart
Description
Restarts the BER test.
Remarks
This command is valide only when the BER Test measurement function is selected.
:INITiate:CONTinuous
Syntax
:INITiate:CONTinuous OFF|ON|0|1
:INITiate:CONTinuous?
Description
Selects continuous (ON|1) or single (OFF|0) sweep mode.
Queries the current sweep mode.
Parameter
Name
Type
Range
Default
--
Bool
OFF|ON|0|1
ON|1
Return Format
The query returns 0 or 1.
Example
The following command sets the instrument to sweep continuously.
:INITiate:CONTinuous ON or :INITiate:CONTinuous 1
The following query returns 1.
:INITiate:CONTinuous?