Rigol RP1018H User Manual Download Page 3

-6- 

1.

 

Specification

 

Model

RP1018H

Division Ratio 

1000:1 

Input Resistance 

200MΩ 

Input Capacitor 

1.5 PF 

Max. Input: DC+AC peak 

18kV CAT II 

Max. Input: AC RMS. 

12kV CAT II 

Max. Loading Current 

90uA 

Bandwidth 

DC~150MHz 

Rise Time 

2.4 ns 

Signal / Noise 

> 60dB at 1kHz ; > 50dB at 1MHz 

DCV Accuracy 

≤ 3% Full Range 

ACV Accuracy 

≤ 3% at 1 kHz 

Temp. Coefficient 

≤ 200 PPM / 

 

Compensation Range 

10PF ~ 35PF 

Safety 

Meets EN61010-031 CAT II 

Cable Length 

2M ± 0.2M 

Operation Temperature Range 

-10 ~ 55

 

Humidity 

85% RH or less (at 35

Storage Temperature Range 

-20 ~ 70

 

Color: Handel/Body 

Black / Yellow 

Weight / Volume 

460g /80(W) x 80(H) X 320(L) mm 

 

 

2.

 

Safety Precautions

 

Thank you for purchasing RP1018H high voltage probe. High voltage probe can 
prevent high voltage testers from subjecting to unexpected shocks when doing high 
voltage measurement work. Before using the high voltage probe, the operator must 
read and fully understand the safety rules. 
In addition, this high voltage probe only for the person who are trained, experienced, 
or otherwise qualified to recognize hazardous situations and who trained in the safety 
precautions that necessary to avoid possible injury when using such a device. 

Summary of Contents for RP1018H

Page 1: ...RIGOL User s Guide Publication Number UGE22X00 1110 RP1018H RP1018H High Voltage Probe Sept 2013 2013 RIGOL Technologies Inc All Rights Reserved ...

Page 2: ...1 2 安全须知 1 3 操作说明 2 4 匹配电容调整 2 5 注意事项 3 6 清洁 4 7 电压测试临界曲线 4 8 附件 5 1 Specification 6 2 Safety Precautions 6 3 Operation 7 4 Compensation Adjustment 7 5 Warning 8 6 Cleaning 9 7 Voltage Derating Curve 9 8 Accessories 10 ...

Page 3: ...RH or less at 35 Storage Temperature Range 20 70 Color Handel Body Black Yellow Weight Volume 460g 80 W x 80 H X 320 L mm 2 Safety Precautions Thank you for purchasing RP1018H high voltage probe High voltage probe can prevent high voltage testers from subjecting to unexpected shocks when doing high voltage measurement work Before using the high voltage probe the operator must read and fully unders...

Page 4: ...e kept clean and free of any conductive contamination The probe is for indoor use only 3 Operation 1 Connect the divider probe common lead alligator clip to a good earth ground or reliable ground 2 Connect the BNC connector to the BNC input of your oscilloscope 3 Select the desired range of your oscilloscope NOTE Please turn the high voltage source off before making any connections 4 Compensation ...

Page 5: ...ection may result in personal injury or damage to the probe or oscilloscope You must be sure to connect the ground before the probe tip comes into contacting with the high voltage And the ground connection must not be removed until the probe tip has been taken away from the high voltage source 2 Do not connect the ground clip to the high voltage source or the probe tip to the ground for any reason...

Page 6: ...cotton cloth lightly moistened with alcohol and water Do not allow any portion of the probe to submerge in the liquid Dry the probe thoroughly before measuring voltage Do not place the probe into the corrosive solvents or fumes as these will cause deterioration to the probe and cables 7 Voltage Derating Curve RP1018H 18kV 3MHz 300V 150MHz 100V 1GHz Voltage dc ac peak Frequency ...

Page 7: ...ccessories Test CRT for H V Adjust bar Carrying Box Contact Us If you have any problem or requirement when using our products or this manual please contact RIGOL E mail service rigol com Websites www rigol com ...

Reviews: