Chapter 2 Command Systems
RIGOL
Programming Guide for DSA1000A series
2-41
Function
The query returns 3 values at comma intervals: carrier power, noise
power and carrier-to-noise ratio (dB).
Explanations
The power unit of returned value is the same as the unit of
Y-axis.
The command transmits the data from the latest measurement
to the ouput buffer.
Return Value Such as -6.01,-6.01,1.00
9.
:FETCh:CNRatio:CARRier?
Syntax
:FETCh:CNRatio:CARRier?
Function
The query returns the carrier power.
Explanations
The power unit of returned value is the same as the unit of
Y-axis.
The command transmits the data from the latest measurement
to the ouput buffer.
Return Value Such as -1.01
10.
:FETCh:CNRatio:CNRatio?
Syntax
:FETCh:CNRatio:CNRatio?
Function
The query returns the C/N Ratio in dB.
Explanations The command transmits the data from the latest measurement to
the ouput buffer.
Return Value Such as 8.956909E-02
11.
:FETCh:CNRatio:NOISe?
Syntax
:FETCh:CNRatio:NOISe?
Function
The query returns the noise power.
Explanations
The power unit of returned value is the same as the unit of
Y-axis.
The command transmits the data from the latest measurement
to the ouput buffer.
Return Value Such as -1.01
12.
:FETCh:EBWidth?
Syntax
:FETCh:EBWidth?
Function
The query returns the measurement result of emission bandwidth
in Hz.
Summary of Contents for DSA1000A Series
Page 1: ...RIGOL Programming Guide DSA1000A Series Spectrum Analyzer May 2011 RIGOL Technologies Inc...
Page 2: ......
Page 14: ......
Page 137: ...Chapter 3 Programming Examples RIGOL Programming Guide for DSA1000A series 3 3...
Page 154: ...RIGOL Chapter 3 Programming Examples Programming Guide for DSA1000A series 3 20 6 Exit code...
Page 158: ...RIGOL Chapter 3 Programming Examples Programming Guide for DSA1000A series 3 24...