Chapter 2 Command Systems
RIGOL
Programming Guide for DSA1000A series
2-115
Front Panel
Peak, Peak Table, Peak Sort, Freq/Ampl
5.
:TRACe:MATH:PEAK:TABLe:STATe
Syntax
:TRACe:MATH:PEAK:TABLe:STATe OFF|ON|0|1
:TRACe:MATH:PEAK:TABLe:STATe?
Function
Turns on or off the Peak Table.
The query returns “0” or “1”.
Default
OFF|0
Front Panel
Peak, Peak Table, State, On/Off
6.
:TRACe:MATH:PEAK:THReshold
Syntax
:TRACe:MATH:PEAK:THReshold NORMal|DLMore|DLLess
:TRACe:MATH:PEAK:THReshold?
Function
Sets the readout type of the peak from Peak Table.
The query returns “NORM”, “DLM” or “DLL”.
Explanations
NORMal: Displaying the first ten qualified peaks in the table.
DLMore: Displaying the first ten peaks that not only meet the
parameter conditions but also have an amplitude greater than
the specified Display Line in the table.
DLLess: Displaying the first ten peaks that not only meet the
parameter conditions but also have an amplitude smaller than
the specified Display Line in the table.
Default
NORMal
Front Panel
Peak, Peak Table, Pk Readout, Normal/>DL/<DL
7.
:TRACe:MATH:STATe
Syntax
:TRACe:MATH:STATe OFF|ON|0|1
:TRACe:MATH:STATe?
Function
Turns on or off the math trace.
The query returns “0” or “1”.
Default
OFF|0
Front Panel
Trace, Trace Math, Operate, On/Off
8.
:TRACe:MATH:A
Syntax
:TRACe:MATH:A T1|T2|T3
:TRACe:MATH:A?
Function
Selects a trace as the trace A for the trace math from T1, T2 or T3.
Summary of Contents for DSA1000A Series
Page 1: ...RIGOL Programming Guide DSA1000A Series Spectrum Analyzer May 2011 RIGOL Technologies Inc...
Page 2: ......
Page 14: ......
Page 137: ...Chapter 3 Programming Examples RIGOL Programming Guide for DSA1000A series 3 3...
Page 154: ...RIGOL Chapter 3 Programming Examples Programming Guide for DSA1000A series 3 20 6 Exit code...
Page 158: ...RIGOL Chapter 3 Programming Examples Programming Guide for DSA1000A series 3 24...