RIGOL
Command Systems
Programming Guide for DS1000E, DS1000D Series
2-32
SLOPe Trigger
1.
:TRIGger:SLOPe:TIME
Command Format:
:TRIGger:SLOPe:TIME
<count>
:TRIGger:SLOPe:TIME?
Function Explanation:
The commands set and query the time setting about slope trigger. The range of
<count> is 20ns~10s.
Returned Format:
The query returns the setting value of <count> in s.
Example:
:TRIG:SLOP:TIME
0.01 Setup the slope time is 10ms.
:TRIG:SLOP:TIME?
The query returns 1.000e-02.
2.
:TRIGger:SLOPe:SENSitivity
Command Format:
:TRIGger:SLOPe:SENSitivity <count>
:TRIGger:SLOPe:SENSitivity?
Function Explanation:
The commands set and query the trigger sensitive. The range of <count> is:
0.1div ~1div.
Returned Format:
The query returns the setting value of <count> in div.
Example:
:TRIG:SLOP:SENS 0.5 Setup the trigger sensitive as 0.5div
:TRIG:SLOP:SENS? The query returns 5.00e-01