RIGOL
User‟s Guide for DS1000CA Series
2-84
Track Mode
Press Cursor
→
Mode
→
Track to display the track mode menu.
Figure 2-116 Table 2-77 Track mode menu
Menu
Settings
Comments
Mode
Track
Set Track mode in cursor measurement.
Cursor A
CH1
CH2
None
Set Cursor A in conjunction with CH1,
CH2 or turn off Cursor A.
Cursor B
CH1
CH2
None
Set Cursor B in conjunction with CH1,
CH2 or turn off Cursor B.
CurA
(Cursor A)
Turn the multifunctional Knob (
)
to
move cursor A Horizontally.
CurB
(Cursor B)
Turn the multifunctional knob (
) to
move cursor B horizontally.
In Track mode, the cross cursor is displayed on the waveform under measuring. If
you move the horizontal position of cursor, the horizontal coordinate and vertical
coordinate of current point as well as the increments of them will be displayed
immediately. Note: horizontal coordinate represents time value, vertical coordinate
represents voltage value.
To do Track mode Cursor Measurements, follow these steps:
1.
Select the
Track
mode for cursor measurement by pressing soft button as:
Cursor
→
Mode
→
Track.
2.
Select channel
Source
for Cursor A and Cursor B by pressing soft button as:
Cursor
→
Cursor A or Cursor B
→
CH1, CH2 or None.
3.
Move the cursors to adjust the horizontal positions of the cursors: