Programming Guide for DM3000 Series
2-84
1.
:scan:create
Syntax:
:scan:create
□
<value>
Function:
The command creates the project.
Explanation:
:scan:create
□
<value>
creates the project. The value is the project name. You may
not set the project name, but you must use the command for the initialization. Or else
the system can not runs the latter operation, and returns the wrong sign.
2.
:scan:addtask
Syntax:
:scan:addtask
□
<value1>
,
<value2>
,
…...
,
<value7>
Function:
The command adds the task for the scan project.
Explanations:
:scan:addtask
□
<value>
adds a scan task, the value are the new task’s parameters,
including task number, channel, function, precision, sampling number and delay time.
Their parameter ranges are shown in Table 2- 14. You only can add the task in number
order. If the two task have the same number, the previous setting will be changed.
Table 2- 14 The Task Parameter and Their Range
Parameter
Type
Range
Task Number
Consecutive Integer
0 ~ 99
*
1
Channel
Consecutive Integer
1 ~ 16
Function
Discrete
DCV, ACV, DCI, ACI,
2WR, FREQ, PERI,
CAP, DIODE
*
2
Range
Discrete
0 ~ 6
(0 indicates “Auto”)
*
3
Precision
Discrete
0, 1, 2
Sampling Number
Consecutive Integer
0 ~ 99
*
4
Delay Time
Consecutive Real
Number
The default unit is s