background image

   

Wiz

 

BLOOD GLUCOSE TEST STRIP INSERT 

Test Result

- Blood glucose test results are shown on the meter as mg/dL or mmol/L, depending on the preset of 

your meter.

- If your blood glucose result is unusually high or low, or if you question your results, repeat the test with 

a new test strip. You can also run a Quality Control Test with RIGHTEST Wiz Control Solution to check 
your meter and test strip. If the test result still remains unusually high or low, contact your healthcare 
professional immediately.

- If you are experiencing symptoms that are not consistent with your blood glucose test results and you 

have followed all the instructions in this manual, contact your healthcare professional immediately.

- RIGHTEST Meter displays results between 10 to 600 mg/dL or 0.6 to 33.3 mmol/L. If your test result is 

below 10 mg/dL (0.6 mmol/L) , " Lo " will appear on the screen. Please repeat your test with a new 
strip. If you still get a " Lo " result, you should immediately contact your healthcare professional.

- If your test result is above 600 mg/dL (33.3 mmol/L), " Hi " will appear on the screen. Please repeat 

your test with a new strip. If you still get a " Hi " result, you should contact your healthcare professional 
immediately.

Limitations

- The meter readings of the blood glucose may be significantly lower than " true glucose levels " in the 

hyperglycemic-hyperosmolar state, with or without ketosis. Critically ill patients should not be tested by 
the RIGHTEST System, or tested with extreme caution.

- Caution is advised in the interpretation of glucose values below 50 mg/dL (2.8 mmol/L) or above 250 

mg/dL (13.9 mmol/L). Consult a physician as soon as possible, if values in this range are obtained.

- Healthcare professionals should evaluate their technique and their patients' technique at periodic 

intervals. To accomplish this, it is recommended to compare the meter result with a  laboratory result 
using the same blood sample. A well characterized clinical laboratory method employing hexokinase 
or glucose oxidase should be used as the comparative method.

Precautions

- Check the expiration date printed on the strip vial. Do not use expired test strips.
- Close the vial cap immediately after taking test strip out from the vial.
- Do not perform quality control test with expired control solution.
- Do not bend or twist the test strip. Damage of test strip may cause wrong result.
- Do not reuse test strips and lancets.
- Discard the used disposable lancet and strip into an appropriate puncture-proof or biohazard container.

If RIGHTEST Meter or Test Strips are exposed to temperature environments out of operating 

  range– below 10°C (50°F) or above 40°C (104°F)– please wait 30 minutes before testing again.
- If you want to purchase new control solutions, please contact your authorized Bionime representative.

8) To remove the lancet, pull off the depth adjustable cap of 

lancing device. Without touching the used disposable lancet, 
stick the lancet tip into the protective cover. Hold the release 
button of lancing device in one hand and pull on the plunger 
in the other hand will safely eject the used disposable lancet 
into an appropriate puncture-proof or biohazard container. 

Warning

- Keep the test strips or vial cap away from children. They may cause a choking hazard. If a test strip or 

vial cap is swallowed, contact your physician immediately.

For more information on how to use your meter,

 

lancing device and understand your test results,  

see the User’s Manual.

6) You will see the countdown mode on the screen. After 5 seconds, the test result appears.
7) Remove the test strip from the meter. Please follow the local regulation and discard the used strip 

properly.

6

7

(1)

Expected values 

GLUCOSE LEVEL

From 70 to 99 mg/dL (3.9 to 5.5 mmol/L)

From 100 to 125 mg/dL (5.6 to 6.9 mmol/L)

126 mg/dL (7.0 mmol/L) and above on more than one testing occasion

INDICATION

Normal fasting glucose

Pre-diabetes (Impaired fasting glucose) 

Diabetes

Fasting Blood Glucose

  0.75 

  1.0 

3

μ

L   

μ

L      1.5 

μ

L       2.0 

μ

L         .0 

μ

L

Please take a minimum of 0.75 

μ

L to do the test on 

glucose monitoring system. Blood sample size 
above 3.0 

μ

L might contaminate the meter.

Sample Size Example

Alternative site testing-palm or forearm blood sampling

 

- To perform a test using samples obtained from alternative sites,

 

install the 

clear cap on the lancing device (For more information on how to install, see 
the Instructions for the lancing device).

- To increase the blood flow, massage the puncture area of palm or forearm 

for a few seconds. 

-

 

Immediately after massaging the puncture area, press and hold the lancing 
device with the clear cap against palm or forearm.   

- Then press the release button. 
- Continue holding the lancing device against palm or forearm and gradually 

increase pressure for a few seconds until the blood sample size is sufficient 
(Refer to Instructions for the lancing device).

Releasing 

button 

Pressing for

a few seconds

Intended Use

RIGHTEST Wiz and Wiz Plus Blood Glucose Monitoring Systems are used by individuals with diabetes. 
It's for checking on glucose levels of whole blood from capillary. Capillary blood can be sampled from 
the fingertip, palm and forearm. It's as an aid in management of diabetes at home and clinical sites.
RIGHTEST Wiz Blood Glucose Test Strips were designed for self-testing outside the body (

in vitro 

diagnostic use). 
The RIGHTEST Blood Glucose Monitoring System aids to diabetes control, could be used by healthcare 
professionals in clinical setting, also by people to use at home.
RIGHTEST System tests the capillary whole blood and provides results equivalent to plasma result in 
laboratory test.
RIGHTEST Wiz Blood Glucose Test Strip is designed for use with RIGHTEST Wiz / Wiz Plus Blood 
Glucose Meter to obtain accurate results.

Performing a Test

1

1) Take one test strip from the vial. Re-cap the vial cap 

immediately.

2) Insert the test strip into the test strip port of the meter with 

the indication symbol facing up.

3) While the blood drop symbol is flashing, you are ready to 

apply the blood sample within 2 minutes. 

4) Place the lancing device against the pad of your fingertip and press the release button.
5) Touch and hold the drop to the edge of sample entry until you hear a " beep " (if volume is turned on)  

and the View Window is totally filled with blood. If the View Window is not totally filled with blood or 
the test does not start, please discard the test strip and repeat the test with a new test strip.

2

4

Preparing the Lancing Device

Wash your hands with warm soapy water and dry thoroughly before you start the test.
1) 

Hold the depth adjustable cap in one hand and hold the hub in the other hand. Bend the cap towards 

the down side, until a gap appears between the cap and hub.

2) 

Pull the cap and hub off in opposite directions, remove the cap.

3) Insert a new disposable lancet firmly into lancet carrier.
4) Twist off and set aside the protective cover of the disposable lancet.
5) Replace the depth adjustable cap. 
6) 

The adjustable cap with 7 depth levels allows you to select the depth of penetration by rotating the 

cap until the preferable depth display in the window. Settings are based on skin type            for soft or 

"

"

thin skin;            for average skin;            for thick or calloused skin. 

"

"

"

"

7) Hold the hub in one hand and pull on the plunger in the other hand. The device will be cocked. Release 

the plunger, it will automatically move back to its original position near the hub.

Test Procedure

3

5

7

6

5

4

3

2

1

Palm

Forearm

Reviews: