4
Pipe ID Control System. RF096-35/55-100 Series
RF096-35/55-100 [Revision 1.0.0] 12.12.2018
5. Basic technical data
Parameter
Value
Measured ID range, mm
35...55
ID measurement accuracy, mm
±0.05
Angle scan resolution, points for turnover
3200
Pipe depth, mm
100
Linear translation accuracy, mm
±0.05
Laser sensor linearity, µm
±15
Laser sensor resolution, µm
3
Laser sensor sample frequency, Hz
9400
Light source
red semiconductor laser,
660 nm wavelength
Laser sensor output power, mW
1
Laser safety class
2 (IEC60825-1)
Interface
Ethernet and PROFINET
Power supply, V
220
Measurement time, s
<5
Environmental resistance
Vibration
20 g / 10…1000 Hz, 6 hours for
each of XYZ axes
Shock
30 g / 6 ms
Permissible ambient light, lx
30000
Relative humidity, %
5-95 (no condensation)
Operating ambient temperature, °С
0…+45
Storage temperature, °С
-20…+70
Weight (without cables), gram
6800
Note:
System parameters can be changed for a specific task.
6. Example of item designation when ordering
RF096-Dmin/Dmax-L
Symbol
Description
Dmin/Dmax
ID measurement range, mm.
L
Measurement depth, mm.
Example:
RF096-35/55-100 - Pipe ID Control System, ID measurement range - 35...55 mm, measurement
depth - 100 mm.