Parameters

If the MODE key is actuated and held down while the meter is switched off, the meter will be switched to system mode after 5 sec.
The set parameters are displayed successively with each key depression:

unit -

To select the pressure unit: 

mbar

torr

 or 

hPa

log -

ON

: to store 

intermediate values 

or

 

OFF

: if only minimum and 

maximum values

 are to be stored

th -

adjust of 

rest period

 (x sec.) till value is stored

cal -

only for 

calibration

, press key again

IR

ON

: to activate the 

IR interface

 (optional) or

 

OFF

:  to deactivate the 

IR interface

 (optional)

hour - cont

:  The meter does not switch itself off in 

measuring mode

 or 

xh

: The meter switches itself off in 

measuring mode

 after 

xh

The value set for the given parameter will be displayed if the key is not actuated within 5 sec.
Then the above described alternatives are changed with each key depression.
The meter is switched again to measuring mode after 5 sec. without a key being depressed.
In AUTO-OFF mode (arrow points downwards on the display) the meter is switched off after approx. 20 sec.

Exchanging batteries

The appearance of the battery-symbol in the bottom left hand corner of the display indicates that the battery is exhausted and must be ex-
changed. To ensure measuring accuracy the last measurement should be repeated once again.
A rechargeable battery may be recharged while built in.

Warning

  Be care to charge with correct power supply (9 V direct-current)

Calibration

The meter should be calibrated in our measuring laboratory at the latest every 6 months. For calibration a special pumping unit is necessary.

VACMASTER concise operating instructions

Switch on the meter with the key

Display: absolute pressure mode “

AUTO-OFF

” (meter is switched off after 20 sec. if no key is pressed)

Key

Display: “

P1

” alternates in seconds cycle with 1

st 

stored value (meter is returned to “AUTO-OFF” mode after 5 sec. if no key is pressed)

Key

Display: “

P2

” alternates in seconds cycle with 2

nd

 stored value

Key

Display: “

P3

” alternates in seconds cycle with 3

rd

 stored value

Key

Display: “

P4

” alternates in seconds cycle with 4

th

 stored value

Key

Display: “

P5

” alternates in seconds cycle with 5

th

 stored value

Key

Display: 

Maximum value

Key

Display: 

Minimum value

Key

Display: “

clr

” for deleting stored data

Key

Display: Absolute pressure, “

Memory

” mode

When intermediate value storage is switched on (see 4. Parameters) a maximum of five pressure values below a threshold of 800 mbar which are
maintained unchanged for longer than 3 sec. (within a band of +1.5% of the actual value, but at least +2 mbar) are successively stored. In the
event that a value has been stored then renewed storage is only possible if pressure values are activated outside the band ± 10 mbar of the last
stored value. The stored values can be displayed consecutively by the display.

Press key twice in quick succession to return to AUTO-OFF mode