5010e_513.doc / 22.01.10
7.2.5 Serial
interface
7.2.5 Serial
interface
A PC or an external printer can be connected to Photometer
5010
via the RS 232 serial interface at the back. A
suitable data cable can be supplied (REF 501-002). The connected device must comply with safety standard
EN 60950.
A PC or an external printer can be connected to Photometer
5010
via the RS 232 serial interface at the back. A
suitable data cable can be supplied (REF 501-002). The connected device must comply with safety standard
EN 60950.
The menu offers following functions:
•
Activation and deactivation of EDP
•
Activation of remote control
•
Activation or deactivation of external
printer with serial interface
•
Activation of download function of oper-
ating system
7.2.5.1 EDP ON / OFF
The current status of the EDP (Electronic Data
Processing) interface is indicated in the first
line by OFF or ON.
Change setting by [SELECT]:
•
EDP OFF: no output,
•
EDP ON (CR-LF): system output through
serial port with CR-LF protocol,
•
EDP ON (STX-ETX-BCC): system output
through serial port with STX-ETX-BCC
protocol,
•
EDP ON (CR-LF-LOG): after each
measurement a formatted string will be
output through the serial port. (e.g. see
table. 7.2.5.1)
Save setting permanently by [OK]
Save setting temporarily up to next switch-off
of the device by [EXIT]
U T I L I T I E S S U B M E N U
E D P O N / O F F
R E M O T E M O D E
S E R I A L P R I N T E R
D O W N L O A D E X I T
3 7 . 0 5 C 0 9 / 0 8 / 0 9 1 1 : 3 9
E D P O F F
E X I T O K S E L E C T
Table 7.2.5.1
Serial
number Method
number ID-NO. Sample
number Result User
Date
Time
7000 20
12345
1
15.5
[user name]
09/08/09
09:30
7.2.5.2 REMOTE CONTROL
Activate remote control by [START].
When activated, Photometer
5010
can be
remote-controlled by a PC and a suitable
program.
Deactivate remote control by sipping lever [P].
By [EXIT] the program returns to the
superordinate menu item.
R E M O T E M O D E
E X I T S T A R T
ROBERT RIELE GmbH & Co KG
Page 50
7- UTILITY PROGRAMS
Y PROGRAMS